Agilent Technologies is a global leader in life sciences, diagnostics and applied markets, recognized for uncompromising integrity in all they do. Their mission is to advance quality of life by focusing their expertise in six key markets. With a team of approximately 18,000 employees spread across the world, their global footprint includes sales offices, logistics centers and business facilities in three major geographies. Their worldwide capabilities enable them to deliver high-quality solutions to their valued customers in 110 countries.
© Agilent Technologies, Inc. Reproduced With Permission, Courtesy of Agilent Technologies, Inc.
Plasmid DNA quality control (QC) needs to distinguish the supercoiled form from linear and open circular isoforms. High-resolution electrophoretic analysis can separate these isoforms while providing sizing and purity measurements from the same analysis.
This application note demonstrates plasmid topology assessment across a 2,000–10,000 bp sizing range, including reproducibility across multiple runs and instruments.
Download this application note to explore:
- How supercoiled, linear, and open circular plasmid isoforms can be resolved with baseline separation across different plasmid sizes
- How supercoiled plasmid sizing showed less than 5% error and 5% CV or less across the tested samples
- How purity measurements identified minor plasmid species, with precision of less than 0.35% CV across multiple runs and instruments
For Research Use Only. Not for use in diagnostic procedures.
Technical Overview
Introduction
As the global demand for biotherapeutics surges, driven by the rise of in vitro
transcribed (IVT) mRNA and adeno-associated virus (AAV) vector-based therapies,
plasmid DNA production has expanded dramatically in recent years. As production
scales, the need for high throughput, reliable analytical methods that can support
modern biomanufacturing workflows has grown. Among the critical quality attributes
(CQAs) evaluated during plasmid release is plasmid topology, including the percentage
of DNA that is in the supercoiled isoform, a key indicator of plasmid integrity and
suitability for downstream applications.
The Agilent Plasmid Topology kit (p/n 5191-6627) for the Agilent Fragment Analyzer
systems enables analysis of supercoiled, linear, and open circular plasmid DNA
isoforms. Designed primarily for quality control (QC) and sizing of supercoiled DNA, the
kit is validated for use with the 5200 and 5300 Fragment Analyzer systems using the
12- and 48-capillary arrays, respectively. As part of continued product development, an
updated vent valve is required for this assay and is available for installation on existing
instruments. This new valve maintains performance specifications and is compatible
with current and future chemistries.¹
The Plasmid Topology kit supports a sizing range of 2,000–10,000 bp with
recommended sample concentrations of 0.1–1.0 ng/µL (Table 1). Each run takes
approximately 45 minutes, and sample volume requirements depend on concentration,
with a minimum of 24 µL per sample. This technical overview highlights the sizing
and resolution capabilities of the kit, and provides examples of QC analysis of several
plasmids, including concentration and percent purity calculations.
Plasmid Topology Assessment
Using the Agilent Fragment
Analyzer Systems
Table 1. Specifications of the Agilent Plasmid Topology kit for the Agilent Fragment Analyzer systems*
Analytical Specifications
DNA sizing range 2,000–10,000 bp
Recommended concentration range 0.1–1.0 ng/µL final concentration DNA in 0.3x pDNA Dilution Buffer
Physical Specifications
Total electrophoresis run time 33 cm: 45 minutes
Samples per run 12, 48
Sample volume required Variable, depends on input concentration (24 µL minimum final sample
volume required)
Guaranteed shelf life 4 months
*Instrument with updated Vent Valve (p/n M5301A) is required to run this assay.
2
Supercoiled plasmid sizing
The Plasmid Topology kit can be used to size the supercoiled
monomer for plasmids across the assay range. To demonstrate the functionality of the system, the sizing accuracy
and precision of the supercoiled isoforms of four different
DNA plasmids spanning the sizing range of the kit from 2,000
to10,000 bp were analyzed across multiple replicates and
multiple instruments. Results from the Fragment Analyzer are
automatically generated by the Agilent ProSize data analysis
software following electrophoresis. Representative electropherogram traces and digital gel images are shown in Figure
2.
Across multiple replicates and instruments, the average
measured sizes closely matched expected values for all four
plasmids (Figure 3A). The pUC19 supercoiled plasmid has
an expected size of 2,686 bp and showed an average size of
2,797 bp with the Fragment Analyzer (Figure 2A). PhiX174
RF1, with a known size of 5,386 bp, had an average size of
5,583 bp (Figure 2B). The expected size of M13mp18 is 7,249
bp, and displayed an average size of 7,433 bp (Figure 2C).
Gag-Pol has an expected size of 9,679 bp, and had an average
size of 10,079 bp (Figure 2D). Each sample showed excellent
sizing accuracy, as demonstrated by the correlation between
the measured and expected sizes (Figure 3A), less than 5%
error (Figure 3B) and high reproducibility, with precision of 5
%CV or less (Figure 3C).
Experimental
Commercial plasmid DNA samples were obtained and diluted
to 20–50 ng/µL with 1X TE buffer (Agilent, p/n DNF-495-
0060). Samples tested include: pUC19 (NEB, p/n N3041S),
PhiX174 RF1 (NEB, p/n N3021S), M13mp18 (NEB, p/n
N4018S), and Gag-Pol (pSB211-CAG-GagpolIIB, BIOVECTRA).
Concentrations were measured using a NanoDrop
spectrophotometer and further diluted to 0.1–1 ng/µL with
0.3x Dilution Buffer. Samples were analyzed on the Agilent
5200 and 5300 Fragment Analyzer systems using the Agilent
Plasmid Topology kit (p/n 5191-6627), according to the kit
guide.² Briefly, 24 µL of sample or pDNA Ladder was aliquoted
to each well of a sample plate, and a separate marker plate
contained 30 µL of pDNA Marker per well. All wells were
overlaid with mineral oil and analyzed in multiple replicates. A
key procedural note is that the inlet buffer tray should contain
800 µL of 1x pDNA Inlet Buffer.
Each plasmid was digested following the manufacturer’s
instructions to generate linear and open circular isoforms for
resolution testing. For each reaction, 2 µL of enzyme and 10
µL of buffer were combined with plasmid DNA and brought
to a final volume of 100 µL with nuclease free water. pUC19,
PhiX174 RF1, and Gag-Pol digests used 2 µL of plasmid DNA,
while the M13mp18 digest used 20 µL. Linear isoforms were
generated using EcoRI HF (NEB p/n R3101S) for pUC19,
M13mp18, and Gag-Pol at 37 °C for 60 minutes, or SacII (NEB
p/n R0157S) for PhiX174 RF1 at 37 °C for 12 hours.
Open circular isoforms were generated using Nt.BspQI (NEB
p/n R0644S) for pUC19, PhiX174 RF1, and Gag-Pol at
50 °C for 1 hour, or Nb.BtsI (NEB p/n R0707S) for M13mp18
at 37 °C for 1 hour. Linear digests were heat inactivated at
65 °C for 20 minutes, and open circular digests were
inactivated at 80 °C for 20 minutes.
Results and discussion
Resolution of plasmid isoforms
An important aspect of plasmid QC is the ability to resolve
and detect supercoiled, linear, and open circular forms of the
plasmid. To evaluate the ability of the Fragment Analyzer and
the Plasmid Topology kit to resolve the different isoforms,
multiple plasmids were enzymatically digested to produce
mixtures of supercoiled, linear, and open circular DNA. Analysis
of four plasmid samples across the sizing range of the kit each
demonstrated baseline separation between all three topologies,
as shown in the electropherogram overlay in Figure 1.
Figure 1. The Agilent Plasmid Topology kit for the Agilent Fragment Analyzer
systems resolves all three isoforms of digested plasmid DNA samples of
varying sizes, as demonstrated in the electropherogram overlay of pUC19,
PhiX174 RF1, M13mp18, and Gag-Pol. Fragment order (from left): lower
marker (LM), supercoiled (SC), linear (L), open circular (OC) forms.
3
Figure 2. Plasmid DNA samples were analyzed on the Agilent Fragment Analyzer systems with the Agilent
Plasmid Topology kit across the sizing range of the assay, from 2,000 to 10,000 bp. Shown are electropherogram
images of (A) pUC19, (B) PhiX174 RF1, (C) M13mp18, (D) Gag-Pol, and (E) a digital gel image of each sample with
the ladder. The primary peak or band shown is the supercoiled form, while the smaller peaks and bands indicate
the presence of low levels of supercoiled dimers, linear, and open circular isoforms, respectively.
A) pUC19 B) PhiX174 RF1
C) M13mp18 D) Gag-Pol
E) Digital gel image (SC only)
4
B) Supercoiled plasmid DNA sizing accuracy
C) Supercoiled plasmid DNA sizing precision
Figure 3. Plasmid DNA samples were analyzed on the Agilent Fragment
Analyzer systems with the Agilent Plasmid Topology kit across the sizing
range of the assay, from 2,000 to 10,000 bp. The sizing (A) linearity,
(B) accuracy, and (C) precision of the supercoiled isoform are shown for
each of the four samples tested. N=23
A) Supercoiled plasmid DNA purity
B) Supercoiled plasmid DNA purity precision
Purity assessment
Assessment of plasmid purity is critical for QC workflows. The
high resolution provided by the Plasmid Topology kit enables
analysis of the supercoiled isoform in relation to the total
sample. The ProSize data analysis software automatically
integrates each peak in the sample to determine percent purity
of the supercoiled monomer relative to total DNA.
The four plasmid DNA samples were analyzed across multiple
replicates with the Plasmid Topology kit (Figure 2A–D). The
electropherogram images of pUC19 (Figure 2A), PhiX174 RF1
(Figure 2B), M13mp18 (Figure 2C) each show a primary peak
representative of the supercoiled monomer. Analysis with the
Plasmid Topology kit also enabled the detection of two smaller
peaks, indicative of a supercoiled dimer and the open circular
isoform. Gag-Pol also showed a small amount of linear DNA
(Figure 2D). The purity of each sample was determined by the
percent total of the supercoiled DNA peak. Each of the samples
showed high percent purity that was consistent between runs
and across multiple Fragment Analyzer systems. pUC19 had
an average purity of 90.9%, PhiX174 RF1 94.4%, M13mp18
94.3%, and Gag-Pol 90.9% (Figure 4A). Notably, the precision of
the purity measurements for each sample was less than 0.35%,
indicating excellent reproducibility between runs and across
multiple instruments (Figure 4B).
Figure 4. Four plasmid DNA samples were assessed on the Agilent Fragment
Analyzer systems with the Agilent Plasmid Topology kit. The purity of the
supercoiled isoform was determined using the percent total concentration
reported by the Agilent ProSize data analysis software. (A) Average purity
and (B) precision from multiple runs and instruments are shown (N=15).
Error bars represent standard deviation.
A) Supercoiled plasmid DNA sizing linearity
5
0
10
20
30
40
50
60
70
80
90
100
0.1 ng/uL 1 ng/uL
Purity (% total concentration)
Sample concentration
Purity
SC SC-D OC
A) 0.1 ng/µL
B) 1 ng/µL
C) Purity
Purity across concentration range
The ability to detect and quantify impurities is essential for
accurate assessment of plasmid DNA quality in QC workflows. In addition to the primary supercoiled monomer,
plasmid preparations can contain low levels of supercoiled
dimers, open circular, and linear forms, or other impurities
that may impact product performance. The high resolution of
the Plasmid Topology kit enables reliable detection of these
peaks across the recommended concentration range, allowing quantitative evaluation of minor species while maintaining
accurate measurement of the primary supercoiled peak.
The Plasmid Topology kit has a recommended concentration
range of 0.1 to 1 ng/µL. To demonstrate the ability of the kit to
detect impurities across this range, the PhiX174 RF1 sample
was analyzed across multiple runs and instruments at both
0.1 and 1 ng/µL. Supercoiled dimer (SC-D) and open circular
(OC) peaks were easily detected at both sample dilutions (Figure 5A–B). Additionally, the percentage of the total concentration for each peak was consistent across both dilutions (Figure 5C). The supercoiled isoform displayed an average purity
of 94.5% at both concentrations, with an excellent precision
of less than 1%. The supercoiled dimer was on average 3.5%,
and the open circular 2% of the total across both concentrations tested. While this sample was able to detect impurities
throughout the recommended concentration range, users
should determine the best concentration for their samples to
ensure that the primary peak is at optimal loading conditions
and impurities can be detected.
Figure 5. PhiX174 RF1 was analyzed on the Agilent Fragment Analyzer
systems with the Agilent Plasmid Topology kit across the concentration
range of the kit. The assay easily resolved the supercoiled (SC),
supercoiled-dimer (SC-D), and open circular (OC) peaks at both (A) 0.1 and
(B) 1 ng/µL. (C) The percent total of each of the peaks was consistent across
both sample dilutions (N=15).
6
A)
B)
Reproducibility
To demonstrate the reliability of the Fragment Analyzer and
the Plasmid Topology kit, a 1:1:1 mixture of the PhiX174 RF1
supercoiled, linear, and open circular isoforms was analyzed
across multiple runs. An electropherogram overlay from the
same capillary over eight successive runs shows the consistent migration time, size, and profile of each of the three
isoforms (Figure 6A). The size and purity of the supercoiled
isoforms were further assessed to demonstrate the reproducibility of the kit (Figure 6B). The average peak size was 5,471
bp, with a precision of 0.95 %CV, while the average purity was
30.7% with a precision of 0.30 %CV, highlighting the strong
reproducibility of the system from run to run.
Conclusions
The Agilent Plasmid Topology Kit provides a robust, accurate,
and highly reproducible solution for characterizing plasmid
DNA across key QC parameters and CQAs. When paired with
the Agilent 5200 and 5300 Fragment Analyzer systems, the
kit delivers accurate and precise sizing of the supercoiled
form within the 2,000 to 10,000 bp range, reliable purity
measurements, detection of minute impurities, and clear
resolution of supercoiled, linear, and open circular isoforms.
These capabilities, combined with consistent performance
across instruments and runs, make the kit a powerful tool for
plasmid quality assessment in research, development, and
manufacturing workflows.
References
1.Vent Valve Update in Agilent Fragment Analyzer
Instruments. Agilent Technologies technical overview,
publication number 5994-8790EN, 2026.
2.Agilent Plasmid Topology Kit. Agilent Technologies kit
quick guide, publication number SD-AT000126, 2026.
Figure 6. (A) A 1:1:1 mixture of the PhiX174 RF1 supercoiled, linear, and open
circular isoforms was analyzed across multiple runs (N=8) on the Agilent
Fragment Analyzer systems with the Agilent Plasmid Topology kit. The size
and purity of the three isoforms were determined using the Agilent ProSize
data analysis software. (B) To demonstrate the reproducibility of the assay,
the sizing and purity of the supercoiled fragment over eight subsequent runs
are shown.
www.agilent.com/genomics/plasmid-dna-kit
For Research Use Only. Not for use in diagnostic procedures.
PR7004-1795
This information is subject to change without notice.
© Agilent Technologies, Inc. 2026
Printed in the USA, June 5, 2026
5994-9279EN
Sponsored by
Agilent Technologies
Agilent Technologies is a global leader in life sciences, diagnostics and applied markets, recognized for uncompromising integrity in all they do. Their mission is to advance quality of life by focusing their expertise in six key markets. With a team of approximately 18,000 employees spread across the world, their global footprint includes sales offices, logistics centers and business facilities in three major geographies. Their worldwide capabilities enable them to deliver high-quality solutions to their valued customers in 110 countries.