Meet Regulatory Standards for Impurity Analysis in Aromatic Solvents
App Note / Case Study
Published: July 30, 2026
Credit: iStock.
High-purity monocyclic aromatic hydrocarbons are essential feedstocks for producing plastics, resins, and textiles. Because trace impurities can affect downstream chemical reactions and product quality, laboratories need analytical methods that deliver reliable, standards-compliant results.
This application note demonstrates dual-channel gas chromatography for analyzing impurities in monocyclic aromatic solvents in accordance with ASTM D7504.
Download this application note to learn how:
- Three analytical methods using helium, nitrogen, and hydrogen carrier gases meet ASTM D7504 requirements
- Dual-channel analysis delivers consistent quantitative results while supporting higher laboratory throughput
- Effective carbon number (ECN) normalization streamlines quantitation and report generation across the analytical workflow
Application Note
Energy and Chemicals
Author
Jie Zhang
Agilent Technologies
(Shanghai) Co. Ltd.
Abstract
This study presents the dual-channel Agilent 8860B gas chromatograph (GC)
system for analyzing the impurities of monocyclic aromatic solvents in accordance
with American Society for Testing and Materials (ASTM) D7504. Methodological
validation confirmed that column resolution, sensitivity, and quantitation accuracy
meet standard requirements. Mass concentration precision (RSDs) achieved is
below 1% for most of the impurities. Both channels produced consistent results
within the reproducibility limits of ASTM D7504. This application note demonstrates
three methods that meet ASTM D7504 requirements using helium, nitrogen,
and hydrogen carrier gas. Meanwhile, the normalization quantification based
on compound's effective carbon number can be easily realized using Agilent
OpenLab CDS software. The 8860B GC together with the OpenLab CDS system
provides a complete and verified solution for aromatic impurity analysis, covering
the whole workflow from sample preparation, separation, and detection to final
report generation.
Analysis of Trace Impurities in
Monocyclic Aromatic Solvents
Following ASTM D7504
Dual simultaneous analysis using the
Agilent 8860B gas chromatograph
2
Implementing ASTM D7504 on a dual-channel GC/FID
platform—where two independent analytical channels
configured with identical injectors, columns, and detectors
operate within a single GC oven—provides significant
operational advantages:
1. Enhanced laboratory throughput: By performing two
injections simultaneously, the test lab can double its
sample capacity per instrument, effectively reducing the
cost-per-analysis and maximizing hardware ROI.
2. Method flexibility and redundancy: A dual-channel set
up allows a lab to run different sample types (such as
toluene on the front channel and xylene on the back)
in a single oven cycle. Furthermore, it provides an
immediate "back up" channel to maintain uptime during
routine maintenance.
For the system to be considered reliable, both channels
should meet ASTM D7504 sensitivity, resolution, and
precision requirements. In addition, both channels should
yield consistent quantitative results for the same sample.
The reproducibility limit (R) demonstrated in ASTM D7504
interlaboratory testing will be used to evaluate the result
consistency between the two channels.
ASTM D7504 recommends helium and hydrogen as carrier
gas. In China, major Chinese petrochemical companies often
align their internal quality control standards directly with
ASTM D7504 when testing for export of high-purity chemical
intermediates. Because nitrogen is commonly used in China,
they commonly adapt the ASTM D7504 method to use
nitrogen carrier gas.
This application note demonstrates how dual-channel
analysis on the Agilent 8860B GC, following ASTM D7504,
delivers consistent and high-precision data using three types
of carrier gas.
Introduction
The production of high-purity monocyclic aromatic
hydrocarbons, such as benzene, toluene, ethylbenzene, and
xylenes (BTEX), is a cornerstone of the modern petrochemical
industry. These compounds serve as essential feedstocks
for the synthesis of plastics, resins, and textiles. Because
even trace impurities can adversely affect downstream
chemical reactions and product quality, rigorous analytical
testing is required to ensure these intermediates meet strict
commercial specifications.
ASTM D7504 is the industry-standard test method for
determining the purity of, and trace impurities in, monocyclic
aromatic hydrocarbons using gas chromatography (GC)
with flame ionization detector (FID).1
This method is favored
for its streamlined workflow; it utilizes effective carbon
number (ECN) correction factors for area normalization. This
eliminates the need for labor-intensive, multipoint calibrations
for every individual impurity, allowing laboratories to calculate
purity and quantify trace nonaromatics and aromatics with
high precision.
The application scope of ASTM D7504 covers a wide
array of aromatic liquids, including benzene, toluene,
mixed xylenes, and styrene. To comply with the method's
stringent performance requirements, the GC system
must demonstrate:
– High sensitivity: A limit of quantitation (LOQ) of at least
0.0006% by mass.
– Good resolution: Maintaining a valley-to-peak ratio of no
more than 50% for critical separations (such as p-xylene
and m-xylene).
– Strict precision: Highly reproducible peak areas to ensure
accurate purity calculations.
3
Chemicals and reagents
Twenty-three chemicals were purchased from ANPEL
Laboratory Technologies (Shanghai) Inc. They are common
impurities found in the monocyclic aromatic solvents and
dissolved in n-hexane to ~ 0.1% mass concentration for
identification of each impurity. The impurity mixture was used
for analytical method optimization. For column resolution and
method recovery evaluation, p-xylene check standard number
one is purchased from AccuStandard Inc.
Lab made check standards containing common
contaminants for benzene, toluene, and p-xylene were
prepared according to the sample composition in previous
work for precision test.2
Benzene in p-xylene (0.0006% wt)
was prepared to evaluate system sensitivity. Detailed sample
compositions are shown in Table 3.
Compound
Benzene
Check
Standard
(Mass %)
Toluene
Check
Standard
(Mass %)
p-Xylene
Check
Standard
(Mass %)
p-Xylene
Standard
No. 1
(Mass %)
LOD Test
Standard
(Mass %)
n-C6 0.032 – 0.003 –
n-C9 – – – 0.017
Benzene 99.91 0.016 0.003 0.02 0.0006
Toluene 0.009 99.92 0.034 0.02
1,4-Dioxane 0.002 – – –
Ethylbenzene – 0.027 0.245 0.1
p-Xylene 0.005 0.013 99.23 99.62
m-Xylene 0.005 0.016 0.376 0.101
Cumene 0.002 – – 0.02
o-Xylene 0.004 0.0020 0.09 0.102
Propylbenzene 0.018 0.0030 – –
Butylbenzene 0.009 – 0.009 –
p-Diethylbenzene – – – 0.02
Table 3. Compositions of test standards.
Experimental
An Agilent 8860B GC was configured with dual split/splitless
inlets and dual flame ionization detectors (FIDs). The sample
introduction was made by dual Agilent 7693A automatic
liquid samplers (ALS) with high density turrets (16-vial). The
analytical parameters and consumables are summarized in
Tables 1 and 2. Data acquisition and analysis were conducted
using Agilent OpenLab CDS version 2.8.
ALS and Inlets
Injection Volume 0.6 µL
Mode Split, 100:1
Temperature 260 °C
Carrier Gas He, 2.1 mL/min H2
, 1.7 mL/min N2
, 1.2 mL/min
Oven Temperature
He Method H2
Method N2
Method
Initial Temperature 60 °C 60 °C 60 °C
Initial Hold 10 min 10 min 15 min
Ramp 1 Rate 5 °C/min 5 °C/min 3.5 °C/min
Ramp 1 Set Point 150 °C 150 °C 150 °C
Ramp 1 Hold 0 min 0 min 0 min
Ramp 2 Rate 30 °C/min 30 °C/min 30 °C/min
Ramp 2 Set Point 220 °C 220 °C 220 °C
Ramp 2 Hold 8 min 8 min 8 min
Post Run Temperature 0 °C 0 °C 0 °C
Post Run Time 0 min 0 min 0 min
Detector
Temperature 260 °C
Air 400 mL/min
Hydrogen 30 mL/min
Make-Up (N2
) 25 mL/min
Table 1. The system analytical parameters using three carrier gas.
Configuration
Column Agilent J&W HP-INNOWax, 60 m × 0.32 mm, 0.5 µm
(p/n 19091N-216I)
Consumables
Inlet Septa Agilent nonstick advanced green (p/n 5183-4759)
Inlet liner* Agilent ultra inert, low pressure drop split liner with glass wool
(p/n 5190-2295)
ALS Syringes Agilent Blue Line autosampler syringe, 5 µL, fixed needle,
23-26s/42/cone (p/n G4513-80206)
Column Ferrules Short graphite for 0.32 mm columns, 10/pack (p/n 5080-8853)
* The deactivated split liner (p/n 5183-4647) is verified and also can be used for
this application.
Table 2. Column and consumables.
4
Results and discussion
Method development for helium, nitrogen, and
hydrogen carrier gas
The helium- and hydrogen-based methods utilized the oven
temperature program recommended by ASTM D7504. The
helium column flow rate used in previous work was reused.3
The hydrogen column flow rate was optimized with 19 of
the 23 typical impurities in monocyclic aromatic solvents
separated at baseline level.
The nitrogen method oven temperature program was
developed based on the helium method using method
translator. The average linear velocity of the helium method
is 29 cm/sec. Theoretically, the optimum linear velocity
for nitrogen is between 10 to 15 cm/sec. To balance the
analytical speed and resolution, the nitrogen average linear
velocity started from 20 cm/sec, with the oven ramp rate
around 70% of the helium method as shown in Figure 1.
To apply the nitrogen method to the 23-impurity mixture
for separation check, minor adjustments were made on the
initial temperature program, extending hold up time at 60 °C
from 14.3 to 15 minutes. Additionally, the ramp rate was
increased to 30 °C/min from 150 °C to 220 °C to accelerate
the nontarget contaminant elution. The tweaked method was
then applied to 23 impurity mixture analysis for separation
check. Results showed 21 compounds can be resolved at
baseline level using nitrogen carrier gas.
Figure 1. The nitrogen method was translated from the helium method using
the method translation tool.
5
Figure 2. (A) Chromatograms of the 23-impurity mixture on the front channel using helium, nitrogen, and hydrogen carrier gases. (B) Chromatograms of
23-impurity mixture on the back channel using helium, nitrogen, and hydrogen carrier gases.
2
4
6
8
Response (pA) Response (pA) Response (pA)
2
4
6
8
12345678 910111213141516171819202122232425262728293031323334353637
2
4
6
8
1
2
3
4
5
6 7
8
9 10
11
12
13
14/15
1617
18
19
20 21
22
23
1
2
3
4
11
12
14/15
16 18
20 21 22 23 19 17
9 10 13
8
6 7
5
1
2
3
4
5
6 7
11
14/15
16
18
20 21
22
23 8
9
10 12
13 17 19
Front channel
×101
×101
×101
Response (pA) Response (pA) Response (pA)
×101
×101
×101
Retention time (min)
123456789 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37
Retention time (min)
2
4
6
8
2
4
6
8
1
2
3 4
5
6 7
8
9 10
11
12 13
14/15
16
17
18
19 20
21
22 23
1
2 3 4
1112
14/15
16 18
20 21 22 19 23 17 13
10 9
8
6 7
5
1
2
3
4
5
6
7
11
14/15
16
18
20
21
22
23
8
9 10 12
13
17
19
Back channel
2
4
6
8
Helium
Nitrogen
Hydrogen
Helium
Nitrogen
Hydrogen
B
A
Chromatograms of the 23-impurity mixture on the two
channels using helium, nitrogen, and hydrogen carrier gases
are presented in Figure 2. Their retention time (RT) can be
used for identification of impurities (RT information is shown
in Table 4).
6
Peak
No. Compounds Name
RT-He-front
(min)
RT-He-back
(min)
RT-N2
-front
(min)
RT-N2
-back
(min)
RT-H2
-front
(min)
RT-H2
-back
(min)
1 n-Nonane 6.541 6.318 9.403 9.116 6.222 6.038
2 Benzene 8.146 7.983 11.738 11.545 7.752 7.630
3 Decane 9.634 9.173 13.768 13.161 9.142 8.764
4 Toluene 12.273 11.992 17.756 17.392 11.776 11.543
5 1,4-dioxane 13.193 13.053 19.147 18.990 12.702 12.603
6 Undecane 14.226 13.678 20.554 19.799 13.723 13.233
7 Ethylbenzene 16.179 15.846 23.496 23.060 15.676 15.390
8 p-Xylene 16.519 16.179 23.989 23.542 16.013 15.721
9 m-Xylene 16.799 16.461 24.397 23.955 16.292 16.022
10 Cumene 18.118 17.760 26.315 25.842 17.604 17.293
11 Dodecane 18.482 17.934 26.779 26.030 17.97 17.478
12 o-Xylene 18.632 18.301 27.064 26.635 18.111 17.827
13 Propylbenzene 19.544 19.179 28.38 27.898 19.107 18.698
14 p-Ethyltoluene 20.093 19.724 29.169 28.685 19.56 19.237
15 m-Ethyltoluene 20.18 19.813 29.296 28.813 19.646 19.326
16 t-Butylbenzene 20.607 20.232 29.912 29.418 20.07 19.742
17 sec-Butylbenzene 21.052 20.664 30.549 30.038 20.508 20.167
18 Styrene 21.388 21.097 31.067 30.696 20.848 20.600
19 Tridecane 22.162 21.621 32.102 31.373 21.62 21.133
20 1,3-Diethylbenzene 22.895 22.506 33.205 32.693 22.333 21.989
21 n-Butylbenzene 23.301 22.912 33.788 33.277 22.731 22.389
22 α-Methylstyrene 23.922 23.598 34.701 34.283 23.353 23.069
23 Phenylacetylene 24.861 24.625 36.074 35.783 24.291 24.010
Table 4. Retention times of the 23 impurities on front and back channels using three carrier gases.
7
Column resolution evaluation
p-Xylene check standard number one was diluted to the
concentration of ~ 0.03% m-xylene using highly pure p-xylene
and then separated under the three methods. The separation
of m-xylene and p-xylene was used to evaluate the column
resolution. ASTM D7504 requires that the distance from
the valley to the baseline between two peaks should not be
greater than 50% of the peak height of the lower peak. For the
m-xylene impurity in p-xylene solvent, the ratio of m-xylene
peak height versus valley-to-baseline height should be no
less than 2:1. As shown in the chromatograms (Figure 3),
the selected Agilent J&W HP-INNOWax column provides H/V
ratio bigger than 5:1, exceeding the resolution criterion easily.
System precision
System precision following ASTM D7504 was evaluated using
lab-made benzene, toluene, and p-xylene check standards.
Ten consecutive dual-simultaneous injections of the check
standards were made under each carrier gas. Representative
chromatograms using helium carrier gas on two analytical
channels are shown in Figure 4.
The precision of peak response and the corresponding mass
concentration on both analytical channels using helium
carrier gas are great with area %RSD below 1.0% for all
compounds except for 1,4-dioxane and cumene due to their
lower concentrations. Among all impurities, 1,4-dioxane in
benzene shows the worst response precision (2.9% ~ 5.2%)
on both channels. Its mass concentration was ~ 0.0022%,
meaning 0.11 ng 1,4-dioxane was loaded on column for
detection. The ECN of 1,4-dioxane is 3.07 indicating its
absolute response factor (response/mass) is less than
one‑third of hydrocarbons and monocyclic aromatic solvents.
The tiny amount and low response property determined its
peak was small and the area precision was not as good as
other compounds.
Figure 3. Resolution of p-xylene and m-xylene on dual channels using
three methods.
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
15.2 15.4 15.6 15.8 16.0 16.2 16.4 16.6
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
22.6 22.8 23.0 23.2 23.4 23.6 23.8 24.0 24.2 24.4 24.6
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
0.4
0.8
1.2
1.6
2.0
Response (pA) Response (pA) Response (pA) Response (pA)
15.6 15.8 16.0 16.2 16.4 16.6 16.8 17.0
0.4
0.8
1.2
1.6
2.0
He front
He back
H2
front
H2
back
N2
front
N2
back
p-Xylene
p-Xylene
p-Xylene
p-Xylene
p-Xylene
m-Xylene
m-Xylene
m-Xylene
m-Xylene
m-Xylene
p-Xylene
×101
×101
×101
×101
Response (pA) Response (pA)
×101
×101
Retention time (min)
Retention time (min)
Retention time (min)
8
Figure 4. (A) Benzene, toluene, and p-xylene check standards on front channel using helium carrier gas. (B) Benzene, toluene, and p-xylene check standards on
back channel using helium carrier gas
0.5
0.6
0.7
0.8
0.9
1.0
Response (pA) Response (pA) Response (pA) Response (pA) Response (pA) Response (pA)
0.5
0.6
0.7
0.8
0.9
1.0
3456789 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37
3 456789 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37
0.5
0.6
0.7
0.8
0.9
1.0
1 2
3
6
8
9
7
6
10
11
4
2
2
1
3
3
5
9 10
5 6 7 9
11
7
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1 2
3
6
5
6
7
8
9
10
11
4
2
2
1
3
3
9 10
5 6 7 9
11
7
Benzene check sample
Toluene check sample
p-Xylene check sample
Benzene check sample
Toluene check sample
Helium method (front channel)
Helium method (back channel)
p-Xylene check sample
×101
×101
×101
×101
×101
×101
Retention time (min)
1. n-Hexane
2. Benzene
3. Toluene
4. 1,4-Dioxane
5. Ethylbenzene
6. p-Xylene
7. m-Xylene
8. Cumene
9. o-Xylene
10. Propylbenzene
11. Butylbenzene
A
B
9
The retention time precision of three check standards running
under helium method is shown in Table 6. The RT %RSD on
the front channel ranged from 0.003 to 0.015% and the RT
%RSD on the back channel was from 0.003 to 0.05%.
Benzene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
%Mass %Mass %RSD Area Area %RSD %Mass %Mass %RSD Area Area %RSD Limits(R)
n-C6 0.031794 0.408 36.974 0.402 0.031695 0.355 34.135 0.544 –9.9E-05
Benzene 99.90796 0 127,747.027 0.454 99.913267 0 117,592.951 0.652 0.005307 0.0247
Toluene 0.008925 0.267 11.288 0.546 0.009076 0.346 10.631 0.632 0.000151 0.011
1,4-Dioxane 0.002144 3.349 0.81 3.559 0.00222 5.157 0.777 5.102 7.6E-05
p-Xylene 0.004996 0.26 6.266 0.608 0.005084 0.642 5.943 0.934 8.8E-05
m-Xylene 0.004998 0.313 6.27 0.645 0.005089 0.567 5.912 0.921 9.1E-05
Cumene 0.001592 1.62 1.985 1.73 0.001615 1.391 1.864 1.691 2.3E-05
o-Xylene 0.003677 0.512 4.613 0.737 0.003735 1.01 4.339 1.223 5.8E-05
n-Propylbenzene 0.018387 0.365 22.92 0.656 0.018692 0.47 21.579 0.907 0.000305
n-Butylbenzene 0.009153 0.36 11.353 0.658 0.009249 0.503 10.624 0.951 9.6E-05
Toluene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
%Mass %Mass %RSD Area Area %RSD %Mass %Mass %RSD Area Area %RSD Limits(R)
Benzene 0.015738 0.131 18.12 0.436 0.015988 0.2 18.711 0.362 0.00025 0.0022
Toluene 99.924051 0 113979.546 0.397 99.923686 0 115961.321 0.361 –0.000365 0.0131
Ethylbenzene 0.026454 0.165 29.927 0.398 0.026501 0.136 30.424 0.374 4.7E-05 0.0029
p-Xylene 0.013029 0.22 14.727 0.39 0.01305 0.217 14.982 0.435 2.1E-05 0.0029
m-Xylene 0.01564 0.157 17.679 0.432 0.01567 0.246 17.99 0.41 3E-05 0.0037
o-Xylene 0.002061 0.67 2.33 0.8 0.002074 0.878 2.381 1.037 0.000013 0.002
n-Propylbenzene 0.003027 0.618 3.401 0.745 0.00303 0.844 3.457 0.976 3E-06 0.0028
p-Xylene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
%Mass %Mass %RSD Area Area %RSD %Mass %Mass %RSD Area Area %RSD Limits(R)
n-C6 0.002626 0.697 2.993 0.817 0.002631 0.971 2.818 1.043 5E-06 0.0043
Benzene 0.002712 0.744 3.398 0.725 0.002729 0.453 3.214 0.542 1.7E-05 0.0013
Toluene 0.033612 0.128 41.658 0.428 0.033803 0.181 39.379 0.416 0.000191 0.0027
Ethyl benzene 0.245836 0.025 305.307 0.484 0.246874 0.026 285.238 0.326 0.001038 0.001
p-Xylene 99.229175 0.001 121973.152 0.482 99.22707 0 114646.657 0.342 –0.002105 0.0173
m-Xylene 0.376303 0.123 462.555 0.548 0.377201 0.12 435.817 0.377 0.000898 0.0165
o-Xylene 0.087991 0.042 108.159 0.505 0.088478 0.065 102.228 0.319 0.000487 0.0024
n-Butylbenzene 0.008862 0.182 10.771 0.575 0.008878 0.224 10.143 0.297 1.6E-05 0.0017
Table 5. Impurities response and quantitation precision for benzene, toluene, and p-xylene check standards using helium carrier gas.
Nitrogen and hydrogen methods showed similar precision
results as the helium methods. Chromatograms obtained
using nitrogen and hydrogen carrier gas and the precision
summary are shown in the Appendix.
10
Limit of detection and limit of quantitation
The method sensitivity was evaluated by running ten
consecutive injections of 0.0006% benzene on both channels
under three carrier gases. The LOD (with SNR at 3:1) and
LOQ (SNR at 10:1) were calculated using the average SNR of
benzene (Table 7). The LODs and LOQs achieved are better
than their corresponding requirements of 0.0002 and 0.0006%
in ASTM D7504. The sensitivity on two test channels is very
close to each other under the same method. The quantitation
limit of the helium method is the best due to the sharper
peaks (compared to the nitrogen method) and cleaner
baseline (compared to the hydrogen method).
Carrier Gas He Method N2
Method H2
Method
Channels Front Back Front Back Front Back
Average SNR 43:1 40.2:1 32.0:1 28.9:1 40:1 36:1
Average Area 0.838 0.818 0.722 0.725 0.758 0.717
Area %RSD 2.336 2.623 2.464 4.779 3.973 3.895
LOD (mass %) 0.000042 0.000045 0.000056 0.000062 0.000045 0.00005
LOQ (mass %) 0.00014 0.00015 0.00019 0.00021 0.00015 0.00017
Table 7. LODs and LOQs calculated based on the average signal to noise
ratio of benzene.
Results consistency of dual-simultaneous injection
The quantification consistency of impurities in three
lab-made check standards are compared between two
channels. According to ASTM D7504, the dual-channel result
consistency can be confirmed with a 95% probability of
being correct if their difference is below the reproducibility
(R) thresholds coming from ASTM D7504 ILS results. As
shown in Table 5, the dual-channel quantitation difference
of each probe impurity (column 10) is below the statistical
R value (column 11) in the helium method except for
the ethylbenzene in p-xylene. The ethylbenzene mass
concentration in the lab-made p-xylene check standard is
0.245%, which has no corresponding R value in ASTM D7504
ILS research results. D7504 shows the R value for 0.0085%
ethylbenzene impurity in p-xylene is 0.001%. The higher the
impurity mass concentration, the bigger the R value. Thus, the
R value for 0.245% ethylbenzene is supposed to be several
times of 0.001%. This assumption can be confirmed from
the R values of 0.002% for 0.0265% ethylbenzene in toluene
as shown in ASTM D7504. The quantitation difference for
0.245% ethylbenzene between two channels is 0.0011%,
much less than the R value for 0.0265% ethylbenzene,
indicating ethylbenzene quantitation between two channels is
also consistent.
Benzene Check Standard
Compounds
Front Channel Back Channel
RT RT RSD% RT RT RSD%
n-C6 3.916 0.008 3.884 0.006
Benzene 8.256 0.011 8.099 0.011
Toluene 12.273 0.009 12.010 0.008
1,4-Dioxane 13.198 0.014 13.073 0.010
p-Xylene 16.515 0.005 16.191 0.005
m-Xylene 16.796 0.006 16.473 0.005
Cumene 18.114 0.005 17.769 0.004
o-Xylene 18.627 0.004 18.311 0.004
n-Propylbenzene 19.539 0.004 19.187 0.004
n-Butylbenzene 23.295 0.002 22.917 0.003
Toluene Check Standard
Compounds
Front Channel Back Channel
RT RT RSD% RT RT RSD%
Benzene 8.148 0.009 7.999 0.007
Toluene 12.505 0.010 12.240 0.010
Ethylbenzene 16.177 0.005 15.859 0.005
p-Xylene 16.517 0.004 16.190 0.004
m-Xylene 16.797 0.004 16.472 0.004
o-Xylene 18.628 0.003 18.309 0.005
n-Propylbenzene 19.539 0.003 19.185 0.004
p-Xylene Check Standard
Compounds
Front Channel Back Channel
RT RT RSD% RT RT RSD%
n-C6 3.917 0.010 3.884 0.012
Benzene 8.148 0.015 7.995 0.047
Toluene 12.274 0.009 12.005 0.050
Ethylbenzene 16.212 0.014 15.891 0.037
p-Xylene 16.800 0.012 16.461 0.032
m-Xylene 16.903 0.009 16.568 0.031
o-Xylene 18.638 0.006 18.315 0.026
n-Butylbenzene 23.294 0.005 22.911 0.017
Table 6. Retention time (RT) and RT %RSD of three check standards in the
helium method.
11
The results consistency between two channels were also
evaluated for the nitrogen and hydrogen methods with the
same conclusion as shown in the Appendix Tables A1 and
A2. The excellent quantitation consistency between the front
and back analytical channels is the foundation of applying
dual-channel simultaneous analysis mode to high throughput
sample analysis or using one channel as a back up for
another channel without calibration standards.
Results accuracy
ASTM does not recommend a reference material for
determining the bias in this test method. Thus, a recovery test
was made to evaluate the quantification accuracy using the
described system. Eight impurities of accurate mass were
added to the high pure p-xylene solvent, in which there is
zero or very low level of impurities. Then the added impurities
were measured. The recovery results based on the test
values, ranging from 92 to 105%, were shown in Table 8 and
Figure 5. The lowest recovery in all three methods compared
to other impurities was m-xylene. It is due to the nonbaseline
resolution of m-xylene impurity and p-xylene solvent. Seven
other compounds demonstrated excellent and comparable
recovery performance in three methods. The ppm level
impurities were detected with satisfactory recovery indicating
the 8860B GC system is a reliable platform for accurate
quality control of aromatic solvents impurity.
Compound
Mass %
Added
*Mass % Measured Results Recovery (%)
He N2 H2 He N2 H2
n-C9 0.0048 0.0048 0.0046 0.0045 98.9 96.2 92.9
Benzene 0.0048 0.0050 0.0048 0.0049 103.7 100.1 102.4
Toluene 0.0041 0.0042 0.0042 0.0042 103.9 102.9 103.1
Ethylbenzene 0.0233 0.0240 0.0240 0.0239 103.3 103.1 102.8
m-Xylene 0.0288 0.027 0.027 0.027 92.7 93.8 93.9
Cumene 0.0052 0.005 0.005 0.005 99.0 100.6 98.6
o-Xylene 0.0241 0.024 0.024 0.024 100.0 100.8 100.0
p-Diethylbenzene 0.0052 0.005 0.005 0.005 98.1 100.6 98.1
* Mass % measured is the average of dual channel results under each method.
Table 8. Recovery of added impurities for p-xylene.
Figure 5. Recovery rate plot of added impurities for p-xylene.
75
80
85
90
95
100
105
110
n-C9
Benzene
Toluene
Ethylbenzene
m-Xylene
Cumene
o-Xylene
p-Diethylbenzene
Recovery (%)
Helium method Nitrogen method Hydrogen method
Quantification and reports by OpenLab CDS
ASTM D7504 quantitates the sample impurity by normalizing
the adjusted response of each peak according to its
ECN. OpenLab CDS software can easily implement peak
area adjustment, normalization, and report generation by
processing raw data using the preset method. The key
settings made in data-processing method include:
– Enabling manual factor mode in the calibration table to
use the ECN of each impurity for response adjustment.
– Enabling normalization function in calibration type
setting and select Apply correction factors to
normalized concentrations to implement adjusted
area-based normalization.
– Selecting NormAmount template or self-defined template
for normalization report generation on target signal.
12
With these method settings and linking the method to
acquired data, a normalization report is generated upon the
date processing is completed. A quantification report for
benzene check standard in the front channel is shown in
Figure 6 as an example.
Figure 6. The quantification report for benzene check standard
(front channel).
Conclusion
The Agilent 8860B gas chromatography system configured
with two identical analytical channels are applied for impurity
analysis in monocyclic aromatic solvents using three carrier
gases. Each analytical channel was evaluated according to
the ASTM D7504 method in terms of resolution, sensitivity,
quantitation precision, and recovery with satisfactory results.
The quantitation results between the two channels are
consistent with each other.
– Column resolution compounds pair demonstrated good
separation with H/V no less than 5:1 for 300 ppm m-xylene
and p-xylene solvent.
– The LODs achieved for benzene on each analytical channel
are less than 0.0001%, exceeding the 0.0002% LOD
required by ASTM D7504.
– The response and mass concentration precisions of most
impurities in check standards are below 1%, meeting or
exceeding the precision requirements in D7504.
– Excellent recovery results, ranging from 93 to 105%,
was achieved for added impurities, indicating good
quantification accuracy based on the 8860B GC system.
– The quantitation difference of all probe impurities between
two channels is less than the reproducibility thresholds (R),
demonstrating excellent across channel consistency.
Besides great instrument performance, the Agilent OpenLab
CDS software can realize ECN based normalization for easy
and fast quantitation and report generation.
The 8860B dual channel GC system coupled with the
OpenLab CDS software provides an accurate and reliable
analysis of impurities in monocyclic aromatic solvents with
satisfactory precision. Lab productivity can be improved
significantly compared to the single channel analytical
system.
References
1. ASTM D7504-23. Standard Test Method for Trace
Impurities in Monocyclic Aromatic Hydrocarbons by Gas
Chromatography and Effective Carbon Number. ASTM
International, 2023.
2. Scott, H. The Analysis of Monocyclic Aromatic
Hydrocarbons by ASTM D7504 on the Agilent 8850
GC System. Agilent Technologies application note,
5994‑7409EN, 2024.
3. Zhang, Y. A Unified Method for the Analysis of Monocyclic
Aromatic Solvents Using the Agilent 8860 GC System
and On-Board Data Processing. Agilent Technologies
application note, 5994-1586EN, 2022.
13
Appendix
Figure A1. Benzene, toluene, and p-xylene check standards on front channel using nitrogen carrier gas.
Figure A2. Benzene, toluene, and p-xylene check standards on front channel using hydrogen carrier gas.
4
5
6
7
8
9
Response (pA)
4
5
6
7
8
9
Response (pA)
2468 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48
4
5
6
7
8
9
Response (pA)
1 2
3
6 9
10
11
4
2
2
1
3
3
5
9 10
5 6 7 9
11
7
Benzene check sample
Toluene check sample
p-Xylene check sample
p-Xylene check sample
×101
0.6
0.8
1.0
Response (pA)
2468 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48
2468 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48
345678 91011121314151617181920212223242526272829303132333435363738
345678 9101112 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38
Retention time (min)
Retention time (min)
×101
0.6
0.8
1.0
Response (pA)
345678 91011121314151617181920212223242526272829303132333435363738
×101
0.6
0.8
1.0
Response (pA)
1 2
3
6
8
10
11
4
2
2
1
3
3
9 10
5 67 9
11
7
Benzene check sample
Toluene check sample
1. n-Hexane
2. Benzene
3. Toluene
4. 1,4-Dioxane
5. Ethylbenzene
6. p-Xylene
7. m-Xylene
8. Cumene
9. o-Xylene
10.Propylbenzene
11. Butylbenzene
1. n-Hexane
2. Benzene
3. Toluene
4. 1,4-Dioxane
5. Ethylbenzene
6. p-Xylene
7. m-Xylene
8. Cumene
9. o-Xylene
10.Propylbenzene
11. Butylbenzene
8
6
7
9
5
6
7
14
Benzene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
Mass% Mass% RSD Area Area RSD% Mass% Mass% RSD Area Area RSD% Limits (R)
n-C6 0.030945 0.657 33.124 0.683 0.031025 0.599 32.662 0.684 –8E-05
Benzene 99.913267 0 117,932.271 0.652 99.913796 0 115,654.795 0.699 –0.000529 0.0247
Toluene 0.009157 0.504 10.66 0.89 0.009046 0.414 10.357 0.877 0.000111 0.011
1,4-Dioxane 0.002134 4.828 0.742 5.076 0.002093 3.639 0.716 3.771 4.1E-05
p-Xylene 0.00518 0.476 5.981 0.95 0.005123 0.741 5.818 1.17 5.7E-05
m-Xylene 0.005188 0.791 5.991 1.336 0.005129 0.737 5.824 0.858 0.000059
Cumene 0.001633 0.844 1.874 1.208 0.001639 1.403 1.85 1.441 –6E-06
o-Xylene 0.0038 0.599 4.387 1.028 0.003772 0.759 4.283 1.111 2.8E-05
Propylbenzene 0.019166 0.556 21.992 0.95 0.018928 0.48 21.36 0.938 0.000238
Butylbenzene 0.009529 0.607 10.879 1.006 0.00945 0.724 10.611 1.092 7.9E-05
Toluene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
Mass% Mass% RSD Area Area RSD% Mass% Mass% RSD Area Area RSD% Limits (R)
Benzene 0.015697 0.2 18.12 0.419 0.015303 0.16 17.36 0.501 3.94E-04 0.0022
Toluene 99.923097 0 114,391.735 0.471 99.922945 0 112,292.422 0.524 1.52E-04 0.0131
Ethylbenzene 0.026886 0.108 30.357 0.532 0.027108 0.151 30.172 0.556 –0.000222 0.0029
p-Xylene 0.013231 0.164 14.94 0.561 0.013321 0.243 14.827 0.504 –9E-05 0.0029
m-Xylene 0.015919 0.215 17.974 0.562 0.016052 0.197 17.866 0.42 –0.000133 0.0037
o-Xylene 0.002093 1.467 2.364 1.624 0.002121 1.626 2.361 1.884 –2.8E-05 0.002
n-Propylbenzene 0.003077 0.58 3.453 0.697 0.00315 0.858 3.485 1.155 –7.3E-05 0.0028
p-Xylene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
Mass% Mass% RSD Area Area RSD% Mass% Mass% RSD Area Area RSD% Limits (R)
n-C6 0.002571 0.879 2.718 0.794 0.002583 0.88 2.856 1.241 –1.20E-05 0.0043
Benzene 0.002684 0.51 3.121 0.74 0.00269 0.937 3.272 1.493 –6.00E-06 0.0013
Toluene 0.033585 0.218 38.62 0.851 0.033417 0.122 40.195 0.751 0.000168 0.0027
Ethylbenzene 0.247285 0.025 282.044 0.774 0.246304 0.031 293.827 0.706 0.000981 0.001
p-Xylene 99.238055 0.002 113,187.456 0.791 99.234597 0 118,381.161 0.718 0.003458 0.0173
m-Xylene 0.376102 0.543 428.981 1.243 0.377082 0.101 449.835 0.642 –0.00098 0.0165
o-Xylene 0.088837 0.075 101.324 0.801 0.088327 0.053 105.368 0.682 0.00051 0.0024
n-Butylbenzene 0.00902 0.3 10.173 0.817 0.008986 0.24 10.6 0.913 3.40E-05 0.0017
Table A1. Impurities response precision in benzene, toluene, and p-xylene check standards using nitrogen carrier gas.
www.agilent.com
DE-015090
This information is subject to change without notice.
© Agilent Technologies, Inc. 2026
Printed in the USA, July 3, 2026
5994-9316EN
Benzene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
Mass% Mass% RSD Area Area RSD% Mass% Mass% RSD Area Area RSD% Limits (R)
n-C6 0.02997 0.621 31.931 1.237 0.029843 0.565 30.028 0.628 0.000127
Benzene 99.904905 0.001 117,028.727 0.736 99.907062 0 110,530 0.359 –0.002157 0.0247
Toluene 0.009356 0.394 10.841 0.639 0.009212 0.395 10.081 0.48 0.000144 0.011
1,4-Dioxane 0.002477 2.937 0.847 2.989 0.002246 3.088 0.734 2.94 0.000231
p-Xylene 0.005357 0.69 6.156 0.614 0.005266 0.508 5.716 0.653 9.1E-05
m-Xylene 0.005377 0.617 6.179 0.559 0.005278 0.633 5.728 0.722 9.9E-05
Cumene 0.001722 1.144 1.967 1.295 0.001695 0.861 1.828 0.963 2.7E-05
o-Xylene 0.003959 0.629 4.55 0.679 0.00388 0.52 4.211 0.627 7.9E-05
Propylbenzene 0.019965 0.736 22.799 0.639 0.019564 0.554 21.101 0.668 0.000401
Butylbenzene 0.00996 0.946 11.317 0.96 0.009792 0.539 10.508 0.652 0.000168
Toluene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
Mass% Mass% RSD Area Area RSD% Mass% Mass% RSD Area Area RSD% Limits (R)
Benzene 0.015375 0.74 17.831 1.405 0.015102 0.752 16.678 0.823 2.73E-04 0.0022
Toluene 99.922293 0 114,619.152 0.99 99.9226 0 1,09153 0.481 –3.07E-04 0.0131
Ethylbenzene 0.027282 0.357 31.038 1 0.027268 0.348 29.542 0.644 1.4E-05 0.0029
p-Xylene 0.01345 0.465 15.302 0.961 0.013394 0.336 14.512 0.582 5.6E-05 0.0029
m-Xylene 0.016169 0.341 18.395 1.026 0.016219 0.546 17.571 0.798 –5E-05 0.0037
o-Xylene 0.002212 2.805 2.516 3.015 0.002186 2.964 2.368 2.833 0.000026 0.002
n-Propylbenzene 0.003219 1.264 3.639 1.115 0.003231 3.479 3.401 1.215 –1.2E-05 0.0028
p-Xylene Check Standard
Impurity
Front Channel Back Channel Difference of
Two Channels
Reproducibility
Mass% Mass% RSD Area Area RSD% Mass% Mass% RSD Area Area RSD% Limits (R)
n-C6 0.002232 1.022 2.459 1.129 0.002407 0.489 2.398 1.076 –1.75E-04 0.0043
Benzene 0.002439 0.47 2.954 1.033 0.002565 0.538 2.81 1.195 –1.26E-04 0.0013
Toluene 0.032358 0.218 38.763 0.869 0.032849 0.16 35.597 0.972 –0.000491 0.0027
Ethylbenzene 0.248691 0.016 295.476 0.928 0.247325 0.071 265.823 1.039 0.001366 0.001
p-Xylene 99.225286 0.001 117,891.859 0.933 99.215055 0.006 1,06636 1.088 0.010231 0.0173
m-Xylene 0.378024 0.197 449.142 1.032 0.390048 1.702 419.274 2.615 –0.012024 0.0165
o-Xylene 0.090252 0.075 107.23 1.04 0.08917 0.063 95.839 1.04 0.001082 0.0024
n-Butylbenzene 0.009438 0.153 11.087 0.982 0.009155 0.268 9.73 0.915 2.83E-04 0.0017
Table A2. Impurities response precision in benzene, toluene, and p-xylene check standards using hydrogen carrier gas.
Brought to you by
Download the Application Note for FREE Now!
Information you provide will be shared with the sponsors for this content. Technology Networks or its sponsors may contact you to offer you content or products based on your interest in this topic. You may opt-out at any time.
Experiencing issues viewing the form? Click here to access an alternate version