Analysis of Digoxin and Metildigoxin in Whole Blood Using Solid-Phase Extraction and Liquid Chromatography Tandem Mass Spectrometry

Complete the form below to unlock access to ALL audio articles.
Abstract
Samples were prepared by SPE extraction with Oasis HLB columns. Separation was achieved with an ACQUITY UPLC HSS T3 column (2.1 × 100; 1.8 μm), at 35°C. The mobile phase consisted of acetonitrile (70%) and ammonium formate 5 mM (30%). Total run time was 1.5 min operating at isocratic mode with a flow rate of 0.3 mL/min. Mass spectrometry detection was performed by positive mode electrospray, on the ammonium adducts with two transitions for each analyte and one for the IS (d(3)-digoxin). The method proved to be specific and linear over the range (0.3-10) ng/mL. This technique also showed high sensitivity with a 0.09 ng/mL LOD and a 0.28 ng/mL LOQ, for both substances. Percentage recovery ranged from 83 to 100% for digoxin and from 62 to 94% for metildigoxin. The intra- and interday precision CV were ≤10%.
The article is published online in the International Journal of Analytical Chemistry and is free to access.