Sensitive Semaglutide Quantitation in Plasma
App Note / Case Study
Published: October 29, 2025
Brought to you by
Agilent Technologies
Agilent Technologies is a global leader in life sciences, diagnostics and applied markets, recognized for uncompromising integrity in all they do. Their mission is to advance quality of life by focusing their expertise in six key markets. With a team of approximately 18,000 employees spread across the world, their global footprint includes sales offices, logistics centers and business facilities in three major geographies. Their worldwide capabilities enable them to deliver high-quality solutions to their valued customers in 110 countries.
Credit: iStock
As the popularity of the GLP-1 receptor agonist semaglutide – sold as “Ozempic” and “Wegovy” – continues to grow, sensitive bioanalytical assays are needed to determine pharmacokinetics and toxicokinetics.
Traditionally, the plasma concentration of semaglutide is measured by a ligand-binding assay, which requires time to develop antibodies and can lack specificity and selectivity.
This application note demonstrates an alternative automated LC-MS workflow for sensitive semaglutide quantitation in human plasma.
Download this application note to explore:
- How automated sample cleanup enhances assay sensitivity compared to conventional protein precipitation methods
- The complete workflow parameters that achieve 0.2 ng/mL LLOQ from minimal sample volume
- Validated methodology meeting regulatory criteria, adaptable to multiple GLP-1 analogs
Application Note
Biopharma/Pharma
Authors
Xi Qiu, Steve Murphy, and
David L. Wong
Agilent Technologies, Inc.
Abstract
Sensitive bioanalytical assays to determine pharmacokinetics and toxicokinetics are
crucial for advancing glucagon-like peptide-1 (GLP-1) receptor agonist medications.
This application note presents an automated workflow using LC/MS to quantify
the GLP-1 analog semaglutide in human plasma. Our results indicate that this
automated LC/MS workflow can enhance assay sensitivity by five times compared
to traditional organic solvent precipitation; this assay achieved LLOQ of 0.2 ng/mL
and was linear up to 1,000 ng/mL from 100 μL starting material. This workflow can
be used for quantitative analysis of GLP-1 therapeutics without the need for specific
antibodies. It also provides excellent sensitivity, high specificity, and fast method
development, which all play a crucial role in drug discovery and development.
Sensitive Quantitation of
Glucagon‑Like Peptide-1 (GLP-1)
Analog Semaglutide from Plasma
Comprehensive workflow for automated sample
cleanup and sensitive quantitation
2
Introduction
In recent years, glucagon-like peptide-1 (GLP-1) receptor agonists have gained significant attention because of their effectiveness to treat obesity and type 2 diabetes. The global market for GLP-1 analogs is projected to grow rapidly, reaching $140 billion by 2030. Among the notable advancements is the peptide semaglutide, which is sold under the brand name "Wegovy" for weight loss and "Ozempic" or "Rybelsus" for diabetes.1, 2 Developing sensitive bioanalytical assays to determine pharmacokinetics and toxicokinetics is crucial for advancing these medications.
Traditionally, the plasma concentration of semaglutide is measured by a ligand-binding assay, which requires time to develop antibodies and can lack selectivity and specificity. Liquid chromatography/mass spectrometry (LC/MS), which is used for sequence identification of GLP-1 receptor agonists, has emerged as an alternative methodology to precisely quantify these molecules in biological matrices because of its superior sensitivity, wide dynamic range, and high specificity.
In this application note, we present an automated workflow with LC/MS technologies to quantify GLP-1 analog drugs in human plasma using an Agilent 1290 Infinity II bio LC and an Agilent 6495D triple quadrupole LC/MS (LC/TQ) system (Figure 1).
Figure 1. Agilent AssayMAP Bravo protein sample prep platform, 1290 Infinity II bio LC, and 6495D LC/TQ.
3
Experimental
Materials and methods
Tirzepatide, semaglutide, formic acid (FA), and ammonium formate were purchased from MilliporeSigma (St. Louis, MO, U.S.). Acetonitrile (ACN) and methanol (MeOH) were purchased from Honeywell (Charlotte, NC, U.S.). The 96-well LoBind plates and protein LoBind tubes were purchased from Eppendorf (Hauppauge, NY, U.S.). Also used were Agilent AssayMAP RP-S cartridges.
Instrumentation
The following instrumentation was used:
–
Agilent AssayMAP Bravo protein sample prep platform (G5571AA)
–
Agilent 1290 Infinity II bio LC system including:
–
Agilent 1290 Infinity II bio high-speed pump (G7132A)
–
Agilent 1290 Infinity II bio multisampler (G7137A)
–
Agilent 1290 Infinity II thermostatted column compartment (G7116A) equipped with a standard flow Quick Connect bio heat exchanger (G7116-60071)
–
Agilent 6495D LC/TQ (G6495DA)
Software
The following software was used:
–
Agilent MassHunter Acquisition software (version 12.0)
–
Agilent MassHunter Quantitative Analysis software (version 12.0)
Sample preparation
Protein precipitation: A mixture of 150 μL ACN and 150 μL MeOH was added to an aliquot of 100 μL human plasma fortified with different concentration of semaglutide. Also, 5 μL of 1,000 ng/mL tirzepatide was used as the internal standard. The mixture was vortexed for 2 minutes and then spun down at 17,000 g for 10 minutes. A 350 μL volume of each supernatant was transferred to a 96-well plate and dried down under nitrogen gas with heating. Then, 110 μL water was added to each well to reconstitute. The solution was further purified using AssayMAP RP-S cartridges.
AssayMAP automated purification: AssayMAP RP-S cartridges were conditioned with 80% ACN and equilibrated with water. Then, 100 μL of the previously mentioned reconstituted samples were loaded onto AssayMAP RP-S cartridges at 3 μL/min, and cartridges were then washed once with water and once with 10% ACN in 10 mM ammonium formate buffer. The final elution step was carried out by loading 30 μL of 5% FA in 80% ACN to the cartridge at 3 μL/min. Lastly, 30 μL water with 5% FA was added to the final elution and 5 μL was injected into LC/MS for analysis.
LC/MS analysis
Data acquisition was performed using an Agilent 1290 Infinity II bio UHPLC coupled to an Agilent 6495D LC/TQ with the Agilent Jet Stream Electrospray ion source. Separation was obtained with an Agilent AdvanceBio Peptide Mapping column (2.1 × 150 mm, 120 Å, 2.7 μm). Tables 1 and 2 list the LC and MS parameters used for this workflow. Positive electrospray ionization of semaglutide yielded [M + 4H]4+ signal at m/z 1,029.4 as the most intense ion. MRM transitions were optimized and 1,029.4 & 1,238.1 was chosen as the quantifier and 1,029.4 & 136.1 was chosen as the qualifier, and the tirzepatide MRM transition (1,204.4 & 396.2) was chosen. Both peptides' MRM transitions are listed in Table 3 with optimal collision energies.
Parameter
Value
Column
Agilent AdvanceBio Peptide Mapping 120 Å,
2.1 × 150 mm, 2.7 μM (p/n 653750-902)
Column Temperature
80 °C
Injection Volume
5 μL
Autosampler Temperature
4 °C
Needle Wash
3 s in wash port (50:50 water:methanol)
Mobile Phase
A) water + 0.1% FA
B) ACN + 0.1% FA
Flow Rate
0.4 mL/min
Gradient Program
Time (min) %B
0 30
3.0 45
5.0 60
5.1 95
8.0 95
8.2 30
10.0 30
Stop Time
10.0 min
Table 1. LC parameters.
4
Parameter
Value
Ion Mode
Positive
Gas Temperature
270 °C
Drying Gas Flow
13 L/min
Nebulizer Gas
25 psi
Sheath Gas Temperature
200 °C
Sheath Gas Flow
12 L/min
Capillary Voltage
5,500 V
Nozzle Voltage
400 V
Table 2. MS acquisition parameters.
Peptide
Precursor Ion
Product Ion
Collision Energy
Tirzepatide
1,204.4
396.2
26
Semaglutide
1,029.4
136.1
42
Semaglutide
1,029.4
1,238.1
30
Table 3. Semaglutide and tirzepatide MRM transitions.
Data processing
All MS data were processed using Agilent MassHunter Quantitative Analysis software.
Results and discussion
Method optimization for GLP-1 analog peptide quantitative analysis
To improve the sensitivity and reproducibility for GLP-1 peptide quantitative analysis, all the sample preparation steps were evaluated to achieve the best results.
1.
Organic solvent composition to precipitate proteins was evaluated, including just ACN, just MeOH, and ACN:MeOH (1:1). The ACN:MeOH (1:1) was found to be the best for semaglutide and tirzepatide extraction (Figure 2).
2.
Loading buffer was evaluated, including pure water, and water with 0.1% FA and 10%, 20%, or 40% ACN. Pure water was found to be the best loading buffer for GLP-1 peptides.
3.
Washing buffer was evaluated, including pure water, water with 0.1% FA, water with 1% FA, and water with 10 mM NH4COOH pH 10 buffer with or without 10% ACN. The combination of water and 10% ACN in 10 mM NH4COOH pH 10 buffer was found to be the best washing solution.
4.
Elution buffers were evaluated, including 10 mM NH4COOH pH 10 buffer with 30% or 80% ACN, and 80% ACN with 0.1%, 1%, and 5% FA. The 5% FA in 80% ACN produced the best results.
Figure 2. Comparison of organic solvents. Left side is tirzepatide MRM (1,204.4 & 396.2) response after organic solvent precipitation, and right side is semaglutide MRM (1,029.4 & 1238.1) response after organic solvent precipitation.
×1
0301235,370.30×10300.51.01.52,706.69×10301235,362.57Acquisition time (min)1.52.02.53.03.54.04.5Counts×10202461,280.68×102012412.89×1020241,055.21Acquisition time (min)1.52.02.53.03.54.04.5CountsTirzepatideSemaglutideACN:MeOH (1:1)ACNMeOHACN:MeOH (1:1)ACNMeOH
5
LC and MS conditions were all optimized, and the peptides' MRM transitions and collision energies were also optimized to achieve the best MS sensitivity for peptide quantitative analysis. The optimized source parameters are listed in Table 2.
Quantitative analysis of semaglutide in human plasma
MassHunter Quantitative Analysis software was used to perform quantitative analysis of calibration curve and QC samples. Overlap chromatograms of blank human plasma and semaglutide low limit of quantification (LLOQ) of 0.2 ng/mL in human plasma are shown in Figure 3A. Figure 3B shows the overlap chromatograms of protein precipitation alone and further purification using AssayMAP RP-S cartridges, which shows that the AssayMAP RP-S cartridges increased assay sensitivity by five fold. The calibration curve was linear up to 1,000 ng/mL with a linear fit and 1/x2 weight, as shown in Figure 3C. Table 3 shows that all calibration points' precision and accuracy are with ± 15% acceptance criteria, demonstrating excellent assay performance.
The intra- and interday analytical precision and accuracy of QC samples were determined from three independent runs performed over three days. The precision and accuracy results for semaglutide in human plasma are shown in Table 4. All levels of QC samples (n = 4) met acceptance criteria of 15%, as recommended by the regulatory agency. The results demonstrated excellent assay performance using automated sample preparation.
Figure 3. (A) Overlap semaglutide MRM (1,029.4 & 1,238.1) chromatograms of blank human plasma and lowest calibration point. (B) Overlap semaglutide MRM (1,029.4 & 1,238.1) chromatograms of protein precipitation and Agilent AssayMAP RP-S cartridge purification for the same concentration of semaglutide. (C) Calibration curve of semaglutide in human plasma from 0.2 to 1,000 ng/mL. The insert is a zoomed-in snapshot of the lowest four calibration points.
0
Acquisition time (min)3.63.83.73.94.04.24.14.34.44.54.64.74.8×102CountsAB050–501001502002503003504004505005506006507007508008509009501,0501,000y = 0.044417x + 7.568103E-004 R2 = 0.9946R = 0.9982Relative responseRelative concentration×101012345600.250.500.751.001.251.501.75Relative responseRelative concentration×10–1C0.2 ng/mL semaglutidein human plasmaBlank human plasmaAssayMAP RP-SProtein precipitation0.10.20.30.40.51.11.21.31.41.50.60.70.80.91.00Acquisition time (min)3.03.54.04.55.05.5×106Counts2.62.82.22.02.41.61.81.21.01.40.60.80.20.43.03.23.40.501.01.52.02.53.03.54.04.5
www.agilent.com
For Research Use Only. Not for use in diagnostic procedures.
RA250627.418
This information is subject to change without notice.
© Agilent Technologies, Inc. 2025
Printed in the USA, August 5, 2025
5994-8530EN
QC Concentration (ng/mL)
0.2
(LLOQ)
0.5
(Low QC)
10.00
(Mid QC)
800
(High QC)
Run 1
Mean
0.197
0.486
8.813
809.632
% Bias
–1.3
–2.8
–11.9
1.2
% CV
6.8
4.5
1.7
2.1
Run 2
Mean
0.206
0.484
10.358
821.419
% Bias
3.1
–3.2
3.6
2.7
% CV
2.3
5.2
3.0
7.2
Run 3
Mean
0.192
0.520
9.378
871.725
% Bias
–4.2
3.9
–6.2
9.0
% CV
7.1
7.3
9.7
3.7
Interday
Mean
0.198
0.497
9.516
834.259
% Bias
–0.8
–0.7
–4.8
4.3
% CV
6.7
6.3
8.8
5.5
Table 5. Semaglutide QC samples' (n = 4) precision and accuracy over three runs.
Conclusion
The developed GLP-1 peptide quantification workflow, combining automation, the Agilent 1290 Infinity II bio LC, and Agilent 6495D LC/TQ system, is an ideal platform for GLP-1 peptide quantitative analysis in biological matrices. Agilent AssayMAP RP-S automated sample cleanup was shown to increase assay sensitivity by five times compared to organic solvent protein precipitation. The developed workflow for semaglutide quantification in human plasma achieved LLOQ of 0.2 ng/mL and was linear up to 1,000 ng/mL from 100 μL starting material, covering a dynamic range of 3.7 order of magnitude. These results are similar or better than reports in the literature. In three individual qualification runs, intra- and interday QC samples' precision and accuracy all met regulatory acceptance criteria, demonstrating excellent assay performance and reproductivity. Another advantage of this workflow is that it is universal and can be applied to many other GLP-1 analogs. It also requires minimal method development time to greatly help drug discovery and development.
References
1.
Ozempic- semaglutide injection, solution. DailyMed. 5 June 2021.
2.
Rybelsus- oral semaglutide tablet. DailyMed. 5 June 2021.
3.
Wegovy- semaglutide injection, solution. 14 Dec. 2021
Concentration (ng/mL)
0.2
0.3
1
5
20
50
200
400
1,000
Run 1
0.223
0.303
0.864
4.250
20.695
49.013
206.798
424.726
985.505
Run 2
0.186
0.279
0.975
4.385
19.75
51.588
216.095
444.826
967.528
Run 3
0.201
0.327
0.916
4.976
18.69
50.559
213.597
434.43
995.441
Mean
0.2034
0.303
0.918
4.537
19.712
50.387
212.163
434.661
982.825
% Bias
1.71
1.00
–8.17
–9.26
–1.44
0.77
6.08
8.67
–1.72
% CV
9.21
7.92
6.05
8.51
5.09
2.57
2.27
2.31
1.44
Table 4. Semaglutide calibration curve (n = 3) results summary.