Streamline Multi-Targeted Steroidomics Workflows With Semi-Automated Sample Preparation
App Note / Case Study
Published: September 17, 2026
Brought to you by
PAL System
PAL System offers integrated automated sample preparation workflow solutions for applications in food safety, environmental science, life science and clinical labs. Solutions are developed and brought to market together with leading analytical instrument manufacturers as well as specialized solution providers.
Credit: PAL SYSTEM.
Plasma steroid analysis is central to clinical research, metabolomics, and toxicology studies, but complex matrices, low analyte concentrations, and labor-intensive sample preparation can limit throughput and reproducibility.
This application note explores the development of a semi-automated sample preparation workflow for steroidomics, enabling the reliable analysis of 47 endogenous steroid compounds in plasma.
Download this application note to discover:
- How semi-automated sample preparation supports the reproducible analysis of endogenous steroids in plasma with excellent repeatability and low contamination levels
- Strategies to reduce operator intervention while maintaining throughput and improving workflow standardization for large-scale studies
- How automation can reduce plastic waste, improve sustainability metrics, and support greener analytical workflows
LC/MS Application Note
Development of a semi-automated sample preparation
protocol for multi-targeted steroidomics in plasma2 LC/MS App Note
Development of a semi-automated sample preparation protocol for
multi-targeted steroidomics in plasma
Authors
Maria Chiara Frondaroli1,2,3, Oriane Strassel1,2,4, Renzo Picenoni5, Hagen M. Gegner5,
Chiara Fanali3, Isabel Meister1,2,4, Serge Rudaz1,2,4
1 School of Pharmaceutical Sciences, University of Geneva, CMU – Rue Michel-Servet 1, Geneva, Switzerland
2 Institute of Pharmaceutical Sciences of Western Switzerland, University of Geneva, CMU – Rue Michel-Servet 1, Geneva, Switzerland
3 Departmental Faculty of Science and Bio-technology, University Campus Bio-Medico of Rome, Via Alvaro del Portillo 21, 00128 Rome, Italy
4 Swiss Centre for Applied Human Toxicology (SCAHT), Basel, Switzerland
5 CTC Analytics AG, Industriestrasse 20, 4222 Zwingen, Switzerland
Keywords
Automation, Micro-SPE,
Steroids
Additional resources
PAL System Content Hub
Micro-SPE
Abstract
Steroids represent a biologically relevant class of compounds, and their quantitative analysis in plasma is essential for investigating
disease mechanisms and metabolic responses. In the context of large-scale cohort studies, the increasing demand for highthroughput and reproducible analytical workflows requires robust and scalable sample preparation strategies.
In this application note, a previously validated extraction procedure was adapted and implemented on an automated PAL
RTC System. The method enabled the reliable analysis of 47 endogenous steroid compounds in plasma samples.
The performance of the semi-automated protocol was evaluated in terms of repeatability, carry-over, and contamination, and
compared with the manual protocol. Automation resulted in improved analytical precision, with reduced variability across
replicates. In addition, the automated workflow significantly decreased the consumption of plastic materials and other laboratory
resources.
Overall, the implementation of the PAL RTC-based semi-automated protocol provides an efficient solution for high-throughput
steroid analysis in plasma, with the added advantage of requiring no direct operator intervention during labor-intensive
operations.www.palsystem.com 3
Introduction
Plasma is a key biofluid for steroid profiling and targeted
metabolomics in clinical and toxicological studies [1] but its
complex composition and the typically low endogenous
concentrations of many steroid species pose significant
analytical challenges. Although steroids share a common core
structure [2], they differ in polarity, degree of conjugation
(e.g., sulfates and glucuronides), and functional groups, which
results in variable extraction efficiencies. In addition, traditional
manual workflows are labor-intensive, time-consuming, and
highly operator-dependent, limiting their suitability for highthroughput applications. For these reasons, sample preparation
remains a critical step.
To address these challenges, there is a growing interest in
automated sample preparation strategies. In this context, the
PAL RTC System offers a robust solution for the standardization
of complex workflows. The PAL RTC can also be implemented
in combination with LC–MS platforms to further extend
automation capabilities, including automated batch or
sequential injection workflows, although it was employed as a
standalone PAL setup in the present study.
In this application note, we present an adaptation and
implementation of a validated steroid procedure [3] onto
a PAL RTC System platform for automated plasma sample
preparation. The workflow was optimized to minimize manual
intervention and ensured consistent processing of 47 steroid
compounds. The semi-automated protocol was evaluated
in terms of contamination, repeatability, carry-over, and
quantitative performances and compared to the conventional
manual protocol. In addition, this application note raises
relevant considerations regarding green analytical chemistry
principles, based on AGREEprep scoring system [5].
Micro-Solid Phase Extraction
Cartridges
Micro-Solid Phase Extraction (µSPE)
is a miniaturized sample preparation
technique designed to clean up or
concentrate analytes using miniaturized
cartridges.
Depending on the sample environment
and solvent strengths, workflows typically
leverage two operational formats:
Pass-Through
Conditions are adjusted so that target
analytes flow straight through the
cartridge without binding. The sorbent
behaves instead as a chemical filter,
actively trapping unwanted matrix
interferences (such as complex lipids or
protein residues).
Bind-Elute
Samples are loaded in an aqueous
environment where target analytes
chemically stick to the sorbent.
Interferences are washed away, and an
organic elution solvent is introduced
later to release and collect the targets. .
Standalone vs. Online PAL RTC Capabilities
While this study demonstrates the utility of the 1.6m PAL RTC as a
standalone workstation dedicated to high-precision liquid handling and
sample preparation, the platform is designed with seamless scalability in
mind.
The PAL RTC can be directly integrated and hyphenated online with
various LC-MS/MS and GC-MS/MS chromatographic platforms. By
bridging the standalone sample preparation sequence directly to an active
injection valve (using the Valve Drive Module), the system transitions into
an automated, “just-in-time” sample prep and injection platform.4 LC/MS App Note
Figure 1: Configuration of the PAL RTC System used to perform the steroid sample preparation workflow with a depiction of
the actual workflow steps. For more details, see 2.2
Sorbent Selectivity: The Role of HLB
The choice of sorbent functionalization determines what is
retained or filtered. For instance, Weak Anion Exchange (WAX)
phases are chemically tailored with ionic groups to target strong
acids like PFAS. For broad biological profiling, HydrophilicLipophilic Balanced (HLB) sorbents are the standard baseline.
HLB is a water-wettable copolymer combining polar and nonpolar building blocks.
Because steroids are neutral, lipophilic molecules when
analyzed as intact Phase II metabolites, literature standardly
utilizes HLB in Bind-Elute mode to retain and concentrate
them from aqueous samples, such as wastewater surveillance
networks.
Materials and methods
Reagents, consumables, equipment and methods
All the solvents, consumables and equipment used are listed
in Table A1. Stable isotope-labeled (SIL) internal standards and
the endogenous steroid ones were purchased from Sigma–
Aldrich (Merck KGaA) and Cambridge Isotope Laboratories
(CIL Inc.).
LC–MS/MS conditions are summarized in Table A2. The
software tools used for data acquisition and data processing
are the same reported in Galmiche et al. [3], where additional
details are also provided. Figure 1 summarizes the configuration
of the PAL RTC System used.
Protocol workflow
The procedural workflow is summarized in Table 1 and
detailed in the steps below.www.palsystem.com 5
In our workflow, we modified this dynamic. The plasma
sample is already extracted using an organic solvent mix
during the protein precipitation step (resulting in a 75%
organic supernatant). This high solvent strength prevents the
target steroids from binding to the non-polar sites of the HLB
phase. Instead, the steroids remain highly soluble and pass
directly into the collection vial, while the cartridge acts as an
inline cleanup filter by permanently trapping residual, highly
hydrophobic matrix interferences. This enables effortless
filtration and high recoveries across a diverse panel of across a
diverse panel of 47 endogenous steroid compounds.
Step 1: Transfer of plasma
Pooled plasma samples are thawed overnight at 4 °C
and subsequently vortexed for 10 s. Aliquots of 250 µl are
transferred into individual vials, which are then closed with
magnetic caps. The vials are placed in the VT54 rack placed in
Tray Holder 1 – Slot 1 to start the semi-automated protocol.
Step 2: Protein precipitation and vortexing
A “SIL mix” containing SIL standards is prepared in a protein
precipitation solution consisting of acetonitrile and MeOH
(9:1, v/v) + 0.1% formic acid. The SIL mix is transferred into
a 50 ml glass bottle connected to port 3 of the dilutor valve.
Using the dilutor, 750 µl of solution are dispensed into each
Step Description
1. Transfer of plasma Vortexing of thawed plasma samples for 10 s and addition of
250 µl plasma to a 2 ml vial
2. Protein precipitation and vortexing Addition of 750 µl of “SIL mix” and vortexing for 10 s
3. Shaking and centrifugation Shaking in the thermomixer at 750 rpm for 15 min at 4 °C
Centrifugation samples at 13’900 g for 15 min
4. µSPE Recovery of the supernatants and filtration through HLB
30 mg µSPE cartridges
5. Liquid handling Transfer of 800 µl from the protein-precipitated filtered samples into 2 ml vials
6. Evaporation Evaporation of samples to dryness in the SpeedVac
7. Reconstitution and vortexing Addition of 50 µl reconstitution solvent, i.e. MeOH/H2O
(50:50) and vortexing for 10 s
8. Shaking and centrifugation Shaking in the thermomixer at 750 rpm for 15 min at 4 °C
Centrifugation samples at 13’900 g for 15 min
9. Liquid handling Transfer of 45 µl of the supernatant to a LC vial
Table 1: Schematic representation of the protocol workflow. PAL RTC System automated actions are indicated in white; procedures carried
out offline/outside the instrument are reported in grey.
vial. Upon completion of this step, each vial is automatically
transported to the vortex module using the tool, vortexed for
10 s, and returned to its original position.
Step 3: Shaking and centrifugation
The vials are shaken at 750 g for 15 min at 4 °C in the
Thermomixer and centrifuged at 13’900 g for 15 min using the
Vial Centrifuge™ for Autosampler Vials MICROSOLV to ensure
efficient protein precipitation.
Steps 4-5: µSPE and liquid handling
The vials are positioned in a VT54 rack located in Tray Holder 2
– Slot 1. Using a 1000 µl syringe mounted on the appropriate
PAL tool, 920 µl of the supernatants are aspirated from the
centrifuged samples, with the needle positioned at a source
penetration depth of 29 mm. To avoid bubble formation, the
aspiration was performed at a controlled rate of 10 µl/s. The
PAL liquid handling system, equipped with the loaded syringe,
then moves to Tray Holder 2 – Slot 3 to pick up the µSPE
cartridge, which is used exclusively for extract filtration. The
assembly is subsequently transferred to Tray Holder 2 – Slot 2,
where the collection vials are located. After each operation, the
syringe is rinsed three times with MeOH delivered by the Fast
Wash station. This procedure is repeated for all the samples. At
the end of the process, 800 µl are transferred from Tray Holder
2 – Slot 2 to a new rack containing collection vials positioned6 LC/MS App Note
in Tray Holder 1 – Slot 1. After each operation, the syringe is
washed with MeOH three times using the Fast Wash station.
Step 6: Evaporation
The samples are dried using a SpeedVac vacuum concentrator.
Step 7: Reconstitution and vortexing
The dried extracts are positioned in Tray Holder 1 – Slot 1 and
reconstituted with 50 µl of solvent (MeOH/H₂O, 50:50) using
the dilutor tool, taking solvent from a glass bottle connected
to the dilutor valve. Once this step is completed for all the
extracts, each vial is vortexed for 10 s.
Step 8: Shaking and centrifugation
The vials are shaken at 750 g for 15 min at 4 °C and centrifuged
at 13’900 g for 15 min.
Step 9: Liquid handling
Using the 100 µl Smart Syringe, 45 µl of the supernatants are
transferred from vials in Tray Holder 1 - Slot 1 to high-insert
vials equipped with blue caps in a Waters rack positioned
in Tray Holder 1 - Slot 3. After each operation, the syringe is
washed with MeOH three times using the Fast Wash station.
The rack is subsequently retrieved and loaded into the LC-MS/
MS autosampler.
Data Analysis
The protocol was evaluated using a set of 16 plasma samples
processed with the PAL RTC semi-automated workflow,
processed in two different batches due to the limited
centrifuge and shaking capabilities. Timing was determined by
recording the total sample preparation time for each workflow.
Repeatability was assessed by calculating the relative standard
deviation (RSD%).
Carry-over was evaluated by injecting blank solvent samples
immediately after a plasma sample and comparing residual
signals to the preceding peak area. Contamination was
assessed through the analysis of water blanks processed
with the same protocol. The performance of the PAL RTC
method was directly compared to the manual protocol [3] in
terms of analytical precision. Quantification was performed
analyzing two different spiking concentration levels with the
internal calibration approach [4]. Trueness and RSDs were
also evaluated. Finally, sustainability was evaluated using the
AGREEprep scoring tool [5].
Results and Discussion
Timings
The parallelization of the protocol for the two processed
8-sample batches is shown in Figure 2. Eight samples
represented the maximum capacity of both the Thermomixer
and the centrifuge.
Without evaporation, the total processing time was
approximately 3 h 30 min, corresponding to about 13 minutes
per sample. When the evaporation step is included, the total
time is about 6 h, increasing the processing time to roughly 23
minutes per sample.
The time needed to process a single sample using the semiautomated workflow was comparable to the one of the
manual protocol. Due to the moderate viscosity of plasma,
samples were aspirated slowly and the syringe was washed
multiple times to prevent carry-over contamination between
consecutive samples. This accounts for the longer time
required for the PAL-assisted µSPE procedure. However,
the µSPE procedure runs without manual intervention from
the operator, resulting in a key advantage of the automated
protocol. An additional consideration is that, although both
semi-automated and manual procedures show similar overall
processing times, automation allows the PAL RTC System to
perform steps that would otherwise require manual handling,
resulting in significant time saving for the operator. Excluding
centrifugation, evaporation, and shaking steps—where operator
involvement is already minimal or absent—the manual
execution of protein precipitation, µSPE, reconstitution, and
transfer can take up to approximately 45 minutes for a batch
of 8 samples. Automating these steps enables the operator to
recover this time and allocate it to other tasks.
RSD, carry-over and contamination
Contamination RSDs and carry-over data are all resumed in
Table A3. Contamination was evaluated across all 16 plasma
samples processed using an identical protocol applied to
water samples.
The ratio between peak areas obtained from processed plasma
samples and those from water blanks indicated contamination
levels ≤ 1% for 34 analytes, between 1% and 2% for 8
analytes, and > 2% for 5 analytes. It should be noted that the
latter group corresponds to low-intensity analytes, for which
measurements are more susceptible to background noise.
RSD values were assessed on a total of 16 samples processed
across two independent batches.
Of the 47 analytes evaluated, 15 exhibited RSDs ≤ 5%, 27
higher than 5 % but still below 10 %, and 5 analytes showed
RSD values between 10% and 12%, indicating excellentwww.palsystem.com 7
Figure 2: Parallelization of the protocol for the two processed 8 sample batches.
analytical repeatability. In terms of carry-over, 37 analytes
displayed values ≤ 1%, 6 analytes between 1% and 2%, and
4 analytes exceeding 2%. As above, the higher values are
associated with low-intensity analytes.
The RSD values used for the comparison between PAL RTCassisted and manual protocols are reported in Table A4, with
a final representation reported in Figure 3. Both protocols
demonstrated good repeatability and yielded comparable
results.
The median RSD values were very similar, at 3.5% for the
PAL-RTC protocol and 3.6% for the manual procedure. The
minimum RSD observed for the PAL-RTC protocol was 0.3%,
compared with 0.7% for the manual method. Conversely, the
maximum RSD values were 13.1% and 15.9% for the PAL-RTC
and manual protocols, respectively.
Figure 3: Comparison between PAL RTC-assisted and manual
protocols in terms of RSD.8 LC/MS App Note
Quantification
Quantification parameters obtained for the PAL-RTC protocol
are reported in Table 2. Using single-point internal calibration,
the method demonstrated good analytical performance in
terms of trueness and RSD. The trueness values ranged from
67 to 122 %, with most values around 100 %, indicating a
good overall agreement between the measured and certified
concentrations. Precision was satisfactory, with RSD values
ranging from 1.0% to 7.4%, except for testosterone (9.6%),
which showed greater variability while remaining within
acceptable limits. Precision at high concentration levels
improved slightly, with RSD values ranging from 2.4% to 4.7%.
Understanding the AGREEprep Metric
Evaluating the “greenness” of sample preparation is essential
for modern sustainable chemistry. The AGREEprep (Analytical
Greenness Metric for Sample Preparation) tool provides a
comprehensive framework to quantify this impact.
Table 2. Trueness and RSD (%) values obtained for the quantification at middle and high spike level.
Middle spike level High spike level
Analyte Certified value
[ng/ml]
Measured value
[ng/ml]
Trueness
(%)
RSD
(%)
Certified value
[ng/ml]
Measured value
[ng/ml]
Trueness
(%)
RSD
(%)
11-Deoxycorticosterone 0.50 0.47±0.01 94 3.1 1.50 1.44±0.05 94 3.6
11-Deoxycortisol 0.50 0.491±0.005 98 1.0 1.50 1.62±0.04 96 2.7
17a-Hydroxypregnenolone 2.33 1.57±0.12 67 7.4 6.98 1.44±0.12 69 2.4
17a-Hydroxyprogesterone 1.26 1.51±0.03 120 1.9 3.77 4.45±0.16 118 3.6
Androstenedione 2.64 2.85±0.05 108 1.9 7.92 8.17±0.38 103 4.6
Cortisol 99.14 101±2 102 1.9 297.42 315±15 106 4.7
Progesterone 0.58 0.58±0.04 101 6.4 1.73 1.41±0.05 105 2.6
Testosterone 0.49 0.60±0.06 122 9.6 1.46 1.82±0.06 111 3.6www.palsystem.com 9
References
[1] N. Ghosh, C. Lejonberg, T. Czuba et al., Analysis of plasma
metabolomes from 11 309 subjects in five population-based
cohorts, Sci. Rep. 14 (2024) 8933.
[2] E. Olesti, J. Boccard, G. Visconti et al., From a single steroid
to the steroidome: Trends and analytical challenges, J. Steroid
Biochem. Mol. Biol. 206 (2021) 105797.
[3] M. Galmiche, O. Strassel, M.A. Monat et al., Extended
Steroid Profiling in Human Serum and Plasma With
Simultaneous Quantitative Determination Using One-Point
Internal Calibration. J. Sep. Sci. 48 (2025) e70147.
[4] G. Visconti, E. Olesti, V. González-Ruiz et al. Internal
calibration as an emerging approach for endogenous analyte
quantification: Application to steroids. Talanta 240 (2022)
123149.
[5] W. Wojnowski, M. Tobiszewski, F. Pena-Pereira, E. Psillakis,
AGREEprep - Analytical Greenness Metric for Sample
Preparation. TrAC 149 (2022) 116553.
Conclusion
The semi-automated sample preparation protocol implemented
on the PAL RTC System proved to be an excellent alternative to
the conventional manual workflow for multi-targeted steroid
analysis in plasma. The method demonstrated good analytical
performance, with good repeatability (RSD ≤ 10% for most
analytes), low carry-over, and minimal contamination.
Although the overall processing time per sample was
comparable to the manual approach, the semi-automated
workflow significantly reduced operator-dependent variability
and improved standardization, which are critical advantages
for routine applications. From a sustainability perspective, the
AGREEprep evaluation highlighted the benefits of automation,
particularly in terms of reduced plastic consumption and waste
generation. In conclusion, the integration of the PAL RTC
System enables a repeatable, efficient, and more sustainable
sample preparation strategy, supporting its application in highthroughput clinical and metabolomics studies.
Sustainability
The environmental sustainability of both the sample
preparation protocols was evaluated using the AGREEprep
tool. The final AGREEprep pictogram consists of a circular
diagram divided into ten sections corresponding to the
principles of Green Sample Preparation, with a color scale
from red to green and a central score summarizing the overall
greenness of the method.
Detailed scoring for both the manual and PAL RTC-assisted
protocols is reported in Tables A5 and A6 and the final
pictograms are shown in Figure 4, with the manual method one
on the left and the semi-automated protocol one on the right.
Both the manual and semi-automated protocols share several
common features. However, the semi-automated method
demonstrates clear advantages in terms of sustainability.
It enables the repeated use of syringes, in contrast to the
disposable plastic tips required in the manual protocol. This
significantly reduces plastic waste generation, although more
organic solvent is used to clean the syringes.
Further improvements in the greenness of the protocol
could be achieved by replacing methanol with ethanol in
the syringe washing step, since ethanol is considered a more
environmentally friendly solvent.
Additionally, the semi-automated approach benefits from a
higher degree of automation, improving workflow efficiency.
The overall assessment indicates that the implementation of
the PAL RTC System results in a markedly more sustainable
sample preparation process.
Figure 4: Comparison of AGREEprep overall scores and pictograms between the manual (on the left) and semi-automated
(on the right) workflows.10 LC/MS App Note
Appendix
Table A1: Consumables and equipment, PAL-RTC System modules and other instruments.
Chemicals and Solvents
Water Fisher Optima 2.5L (Art. No. W6500)
MeOH Fisher Optima 2.5L (Art. No. A4556212)
Acetonitrile Fisher Optima 2.5L (Art. No. A955212)
Ammonium fluoride Aldrich 33869-25G, >99.99%
Formic Acid Biosolve BV, 99%, ULC/MS, 50 ml
Consumables and equipment
Vials TruView pH Control LCMS Certified Clear Glass, 12 x 32 mm, Screw Neck Vial, Max Recovery, with
Cap and preslit PTFE/Silicone Septum, 1.5 ml Volume, 100/pk (Art. No. 186005662CV)
Vials (for injection) TruView pH Control LCMS Certified Clear Glass, 12 x 32 mm, Screw Neck Vial, Total Recovery, with
Cap and preslit PTFE/Silicone Septum, 1.5 ml Volume, 100/pk (Art. No. 186005663CV)
Magnetic caps Schraubverschluss, N 9, PP/Stahl, magnetisch, Loch, Silikon w./PTFE rot, 1,0mm (Art. No. 702155) –
Macherey-Nagel AG
Blue caps
(for injection)
Schraubverschluss, N 9, PP, blau, Loch, Silikon w./PTFE blau, geschlitzt, 1,0mm (Art. No. 702288.1)
– Macherey-Nagel AG
µSPE cartridges µSPE cartridges (HLB – 30 mg) (Lot. No. 734375)
Rack PAL System Rack for 54 Vials of 2 ml (Art. No. C-PAL3-RACK-VT54)
Waters Rack Waters Vial Tray - 48 Well (12 mm x 32 mm Vials) (Art. No. 700011047)
µSPE Kit 1 pc Tray Holder
2 pcs VT54 rack (PAL3-Rack-VT54)
1 pc µSPE Cartridge Tray, 54 position, aluminum
1 pc µSPE Elution Tray Cover, 54 position aluminum
1 pc Waste Receptacle with drain
4 pcs PAL System µSPE Cartridges for initial installation, 3-point teaching
(Art. No. PAL3-HW-uSPE54)
Dilutor kit Dilutor Single, PAL Bus Connection cable, Dilutor Tool, Dilutor Syringe 1 ml, Needle NDL-S 22-3,
PFA Transfer Tube 1/16» (Art. No. PAL3-DilutorSingle)
1000 µl Smart Syringe PAL System Smart Syringe 1000 µl, PTFE Plunger, G 23, Needle Length 57 mm, Point Style Conical,
for Tool D8/57 (Art. No. SF1000-57-T-23-CO)
100 µl Smart Syringe PAL System x-type Smart Syringe 100 µl, G 22S, Needle Length 57 mm, Point Style Flat, for Tool
D7/57 (Art. No. SFX100-57-T-22S-FL)
Liquid Syringe Tool PAL System Liquid Syringe Tool for syringes of 250 µl - 1000 µl with a needle length of 57 mm (Art.
No. PAL3-TH-D857)
Liquid Syringe Tool PAL System Liquid Syringe Tool for 1.2 - 100 µl Syringes, Needle Length 57 mm (Art. No. PAL3-
TH-D757)www.palsystem.com 11
PAL RTC System modules
PAL liquid handling
robot PAL RTC
Cooler Stack Peltier Stack 6DW for storage under defined temperature conditions between 4 °C and 40 °C (Art.
No. PAL3-PeltierStack6DW)
Vortex Mixer Standard vial sizes: 2 ml / 10 ml / 20 ml
1 additional slot for custom specific vials
(Art. No. PAL3-VortexMixer)
Solvent Module Module with 3 x 100 ml solvent container (glass) with septum cap (Art. No. PAL3-SolventModule)
Fast Wash Station 1 pc Wash Module with two wash positions and one waste position
1 pc Pump Module with two solvent pumps
2 pcs 1 Liter Solvent Bottles with tubings and frits
1 pc PAL BusConnection cable
(Art. No. PAL3-WashSolv)
Out of PAL RTC System equipment
Thermomixer Eppendorf Thermomixer® C (Art. No. 5382000023) with adaptor (Art. No. 5309000007)
Centrifuge Vial Centrifuge™ for Autosampler Vials MICROSOLV (ref. 9512S-1DMP-PX, plate rotor CE10-C0817-
VR)
SpeedVac Thermo Savant SC210A (concentrator), RVT400 (cold trap)
Table A2: LC and MS/MS parameters.
LC conditions
LC system Acquity H-Class UHPLC system (Waters Corp.)
Column Restek Raptor Inert Biphenyl (2.1 x 100 mm, 1.8 µm particle size, 100 Å) with precolumn
(2.1 x 5 mm)
Mobile phase A Water + 0.01% formic acid
Mobile phase B MeOH + 0.01% formic acid
Flow rate 0.4 ml/min
Gradient 0-18 min 40-100% B, 18-21 min 100% B, 21-21.2 min 100-40% B, 21.2-46.2 min 40% B
Column temperature 45 °C
Injection volume 5 µl
Autosampler temperature 6 °C
MS/MS conditions
MS system Xevo TQ-XS Triple Quadrupole Mass Spectrometer (Waters Corp.)
Ionization source ZSpray ESI source
Post-column infusion NH4F 5mM in MeOH from 3.5 to 20.0 min at 5µl/min
Capillary voltage 1.7 kV
Source temperature 150 °C
Desolvation temperature 600 °C
Desolvation gas flow 800 L/h
Cone gas flow 150 L/h
Probe position 1–5 marks
Probe angle 45°12 LC/MS App Note
Table A3: Contamination, RSDs and carry-over values.
Compound Name Contamination effect RSDs Carry over
11-hydroxyandrostenedione 0.4% 5.6% 0.5%
11B-hydroxytestosterone 0.9% 7.2% 2.0%
11-dehydrocorticosterone 0.1% 4.1% 0.2%
11-deoxycorticosterone 7.6% 9.9% 5.9%
11-deoxycorticosterone 13C3 0.0% 11.6% 0.0%
11-deoxycortisol 1.1% 4.9% 0.6%
11-deoxycortisol 13C3 0.0% 6.0% 0.0%
11-ketotestosterone 0.7% 7.5% 0.3%
16A-hydroxyprogesterone 0.2% 6.0% 0.4%
17A,20A-dihydroxyprogesterone 1.5% 8.0% 1.2%
17A-hydroxypregnenolone 1.6% 4.4% 0.4%
17A-hydroxypregnenolone 13C2 d2 1.9% 5.2% 7.6%
17A-hydroxyprogesterone 1.7% 4.9% 0.3%
17A-hydroxyprogesterone 13C3 0.3% 7.5% 0.0%
18-hydroxycortisol 0.2% 4.6% 0.1%
20A-cortolone 0.3% 9.2% 0.0%
20A-dihydrocortisol 0.1% 7.8% 0.0%
20A-dihydrocortisone 0.0% 3.7% 0.0%
20A-dihydroprogesterone 0.8% 6.0% 0.4%
20B-cortolone 2.2% 10.1% 1.4%
20B-dihydrocortisol 0.1% 4.5% 0.1%
20B-dihydrocortisone 0.7% 5.1% 0.2%
3A,5A-tetrahydrocortisol 0.0% 6.4% 0.0%
5A/B-dihydrocortisol 0.2% 8.1% 0.0%
5A-dihydrotestosterone 13C3 1.5% 6.5% 0.2%
6A-hydroxytestosterone 1.4% 10.5% 1.7%
Allopregnenolone 4.6% 10.8% 7.5%
Androstenedione 1.2% 7.7% 1.1%
Androstenedione 13C3 0.1% 8.1% 0.2%
Androsterone 3A-Glucuronide 0.0% 3.8% 0.0%
Cholic acid 0.0% 5.9% 0.1%
Corticosterone 0.0% 6.7% 0.1%
Cortisol 0.0% 4.6% 0.0%
Cortisol 13C3 0.0% 6.0% 0.0%
Cortisone 0.0% 3.6% 0.0%
Cortisone 13C3 0.0% 5.0% 0.0%
DHEAS (Dehydroepiandrosterone sulfate)
+ Epitestosterone 17A-Sulfate
0.0% 2.5% 0.0%
DHEAS 0.0% 2.5% 0.0%
Glycochenodeoxycholic acid 0.0% 7.2% 0.0%
Glycoursodeoxycholic acid 0.0% 2.5% 0.0%
Pregnanetriol 3.7% 11.7% 1.3%www.palsystem.com 13
Pregnenolone 13C2 d2 4.6% 8.9% 5.0%
Progesterone 0.0% 6.1% 0.1%
Progesterone 13C3 0.0% 5.9% 0.0%
Testosterone 0.0% 3.8% 0.0%
Testosterone 13C3 0.0% 7.1% 0.0%
Testosterone 17B-Sulfate 0.0% 2.5% 0.0%
Table A4: PAL-RTC vs manual protocols comparison.
Compound Name PAL-RTC RSD Manual RSD
11-hydroxyandrostenedione 1.5% 4.5%
11B-hydroxytestosterone 2.9% 7.0%
11-dehydrocorticosterone 2.1% 1.6%
11-deoxycorticosterone 5.0% 7.0%
11-deoxycorticosterone 13C3 5.7% 6.1%
11-deoxycortisol 6.3% 2.6%
11-deoxycortisol 13C3 2.8% 2.5%
11-ketotestosterone 7.5% 4.4%
16A-hydroxyprogesterone 4.4% 5.2%
17A,20A-dihydroxyprogesterone 4.2% 5.9%
17A-hydroxypregnenolone 7.8% 14.7%
17A-hydroxypregnenolone 13C2 d2 6.8% 6.2%
17A-hydroxyprogesterone 2.3% 6.7%
17A-hydroxyprogesterone 13C3 5.9% 1.2%
18-hydroxycortisol 5.1% 4.9%
20A-cortolone 13.1% 5.0%
20A-dihydrocortisol 7.9% 1.9%
20A-dihydrocortisone 2.2% 3.2%
20A-dihydroprogesterone 1.3% 9.2%
20B-cortolone 2.1% 2.2%
20B-dihydrocortisol 8.9% 4.9%
20B-dihydrocortisone 2.5% 0.7%
3A,5A-tetrahydrocortisol 5.9% 1.8%
5A/B-dihydrocortisol 9.0% 1.9%
5A-dihydrotestosterone 13C3 5.9% 7.5%
6A-hydroxytestosterone 5.9% 15.9%
Allopregnenolone 3.3% 7.3%
Androstenedione 7.4% 1.4%
Androstenedione 13C3 3.9% 7.1%
Androsterone 3A-Glucuronide 5.1% 5.3%
Cholic acid 11.6% 1.4%
Corticosterone 3.2% 2.2%
Cortisol 2.6% 3.6%
Cortisol 13C3 2.1% 2.8%
Cortisone 1.6% 0.8%14 LC/MS App Note
Cortisone 13C3 2.3% 1.4%
DHEAS (Dehydroepiandrosterone sulfate) + Epitestosterone 17A-Sulfate 3.3% 2.9%
DHEAS 1.8% 1.4%
Glycochenodeoxycholic acid 3.1% 2.4%
Glycoursodeoxycholic acid 0.7% 6.0%
Pregnanetriol 2.7% 15.6%
Pregnenolone 13C2 d2 7.3% 9.6%
Progesterone 3.5% 3.0%
Progesterone 13C3 2.9% 1.0%
Testosterone 0.3% 0.7%
Testosterone 13C3 3.8% 4.4%
Testosterone 17B-Sulfate 1.6% 0.8%
Bin PAL-RTC RSD Manual RSD
Analytes in the bin Analytes in the bin
< 1% 2 4
< 3% 19 21
< 6% 36 33
< 9% 44 42
< 12% 46 44
< 15% 47 45
< 18% 47 47
Table A5: Manual protocol scoring details.
# Criterion Selection Score Weight
1 Sample preparation placement On site 0.33 1
2 Hazardous materials 0.75 ml 0.37 5
3 Sustainability, renewability, and
reusability
<25% of reagents and materials are
sustainable or renewable, but can only be
used ONCE
0.00 2
4 Waste 10 g 0.26 4
5 Sample size economy 0.25 g 0.87 2
6 Sample throughput 3 or 4 samples/hour 0.33 3
7 Integration and automation 3 steps, semi-automated systems 0.38 2
8 Energy consumption 10 W 1.00 4
9 Post-sample preparation configuration Advanced MS (ICP-OES, ICP-MS, etc.) 0.00 2
10 Operator safety 2 hazards 0.50 3www.palsystem.com 15
Table A6: Semi-automated protocol scoring details.
# Criterion Selection Score Weight
1 Sample preparation placement On site 0.33 1
2 Hazardous materials 3 ml 0.17 5
3 Sustainability, renewability, and
reusability
>75% of reagents and materials are
sustainable or renewable
0.75 2
4 Waste 5 ml 0.37 4
5 Sample size economy 0.25 g 0.87 2
6 Sample throughput 3 or 4 samples/hour 0.33 3
7 Integration and automation 3 steps, fully automated systems 0.75 2
8 Energy consumption 20 W 0.82 4
9 Post-sample preparation configuration Advanced MS 0.00 2
10 Operator safety 0 hazards 1.00 3Imprint
Date of print: 07.2026
CTC Analytics AG
Industriestrasse 20
CH-4222 Zwingen
Switzerland
T +41 61 765 81 00
Contact: info@ctc.ch
Legal Statements
CTC Analytics AG reserves the right to make improvements and/or changes to the
product(s) described in this document at any time without prior notice.
CTC Analytics AG makes no warranty of any kind pertaining to this product, including
but not limited to implied warranties of merchantability and suitability for a particular
purpose.
Under no circumstances shall CTC Analytics AG be held liable for any coincidental
damage or damages arising as a consequence of or from the use of this document.
© 2023 CTC Analytics AG. All rights reserved. Neither this publication nor any part
hereof may be copied, photocopied, reproduced, translated, distributed or reduced
to electronic medium or machine readable form without the prior written permission
from CTC Analytics AG, except as permitted under copyright laws.
CTC Analytics AG acknowledges all trade names and trademarks used as the property
of their respective owners.
PAL is a registered trademark of CTC Analytics AG | Switzerland
www.palsystem.com
Visit our homepage for more information.
Sponsored by
PAL System
PAL System offers integrated automated sample preparation workflow solutions for applications in food safety, environmental science, life science and clinical labs. Solutions are developed and brought to market together with leading analytical instrument manufacturers as well as specialized solution providers.
Download the Application Note for FREE Now!
Information you provide will be shared with the sponsors for this content. Technology Networks or its sponsors may contact you to offer you content or products based on your interest in this topic. You may opt-out at any time.
Experiencing issues viewing the form? Click here to access an alternate version