Transition to Hydrogen Carrier Gas for Nitrosamine Analysis
App Note / Case Study
Last Updated: July 10, 2024
(+ more)
Published: June 27, 2024
Credit: iStock
Helium scarcity is a growing concern for laboratories, creating demand for hydrogen as an alternative carrier gas for GC-MS analyses.
While hydrogen offers faster run times and better chromatographic resolution, it can react with analytes and suffer from a loss of sensitivity.
This application note explores the suitability of hydrogen carrier gas, in comparison to helium, for the analysis of eight nitrosamine impurities found in pharmaceutical products.
Download this application note to discover:
- How hydrogen carrier gas performs compared to helium in GC/MS analyses
- Strategies to maintain spectral quality and accuracy with hydrogen
- Tips for efficient data management
Application Note
Pharma & Biopharma
Authors
Soma Dasgupta, Vivek Dhyani,
Anastasia Andrianova, and
Joel Ferrer
Agilent Technologies, Inc.
Abstract
In response to helium scarcity, labs are exploring alternative carrier gases for gas
chromatography/mass spectrometry (GC/MS) analyses. This application note
shows the suitability of hydrogen as a carrier gas and compares it with helium for
the analysis of eight nitrosamine impurities in certain sartan drugs using either
an Agilent 8890 gas chromatograph system coupled to an Agilent 7010 Series
triple quadrupole (TQ) GC/MS system with a high-efficiency ion source (HES),
or an Agilent 7000E triple quadrupole GC/MS with an Agilent HydroInert ion
source. The spectral matches against the NIST library for the eight nitrosamine
impurities analyzed with hydrogen carrier gas ranged from 79 to 97. Excellent
calibration linearity was observed over the concentration range of 0.3 to 50 ng/mL
with R2
> 0.99. The sensitivity requirements were met at 0.03 ppm following a
signal‑to-noise ratio (S/N) requirement of 10. The integration of OpenLab ECM XT
with MassHunter Acquisition 13.0 streamlines data management, providing analysts
with centralized access to instrument-generated data, fostering collaboration,
maintaining data integrity, and optimizing workflow processes.
Quantification of Nitrosamine
Impurities in Sartan Drugs Using
an Agilent GC/TQ with Hydrogen
Carrier Gas
2
Introduction
Helium availability has been a concern for several years,
which has led to a significantly increased interest in
transitioning to alternative carrier gases such as hydrogen.
While hydrogen is an efficient GC carrier gas, understanding
its reactivity with analytes is crucial for obtaining accurate
results. Hydrogen has some advantages over helium,
including faster run times and better chromatographic
resolution. However, hydrogen also has some challenges,
such as potential loss of sensitivity and spectral changes
arising from hydrogen reactions with sample analytes. Such
reactions would alter the mass spectrum of a peak in the total
ion chromatogram (TIC), leading to potential misidentification
of compounds. Transitioning from helium to hydrogen carrier
gas is a significant change that requires careful planning and
execution. The "EI GC/MS Instrument Helium to Hydrogen
Carrier Gas Conversion Guide1
" provides comprehensive
instructions to facilitate the transition. Also, the introduction
of the HydroInert source allowed for the preservation
of spectral fidelity even for reactive compounds in the
presence of hydrogen. As a result, various applications were
successfully performed using hydrogen carrier gas, including
the analysis of volatile organic compounds2
, polycyclic
aromatic hydrocarbons (PAHs)3,4, and the target compounds
listed in EPA TO-15.5
Apart from these in-demand applications, another application
of importance is the analysis of nitrosamine impurities in
pharmaceutical products. Using helium as a carrier gas
has been adopted widely for the analysis of nitrosamine
impurities in pharmaceuticals. Nitrosamines may react with
hydrogen under certain conditions, undergoing conversion to
undesirable amines or hydrazines. Therefore, it is important
to establish that spectral quality is not affected when using
hydrogen carrier gas. This application note evaluates the
use of hydrogen carrier gas with the HES and HydroInert
ion sources as part of a method for the analysis of eight
nitrosamine impurities: nitrosodimethylamine (NDMA),
N-nitrosomethylethylamine (NMEA), N-nitrosodiethylamine
(NDEA), N-nitroso-ethylisopropylamine (NEIPA),
N-nitrosodiisopropylamine (NDIPA), N-Nitrosodipropylamine
(NDPA), N-nitrosodi-n-butylamine (NDBA), and
N-Nitrosopiperidine (NPIP). The results obtained were
evaluated for spectral quality, linearity, repeatability, recovery,
and compliance with current regulations for the analysis of
nitrosamines in pharmaceutical products.
Experimental
The active pharmaceutical ingredients (APIs) and drug
products tested included valsartan, irbesartan, losartan, and
olmesartan. A portion of 500 mg of drug substance was
accurately weighed into a disposable 15 mL glass centrifuge
tube, and 5 mL of internal standard solution (~50 ng/mL
NDMA-d6
in dichloromethane) was added using a volumetric
pipette. The samples were vortexed for 1 minute, then
placed in a centrifuge and spun at 4,000 rpm for 5 minutes.
The undissolved drug substance settled at the bottom.
Using a disposable pipette, approximately 2 mL of the
dichloromethane layer was filtered through a 0.45 µm nylon
filter and transferred to a GC vial for analysis.
Standard preparation
The standard stock was diluted to obtain a calibration
solution in the range of 0.3 to 50 ng/mL and prepared in
dichloromethane containing NDMA-d6
as internal standard.
Instrumentation and analysis
Analyses were performed using an 8890 GC system
equipped with an Agilent 7693A automatic liquid sampler
(ALS) coupled to a 7010 Series GC/TQ with an HES and a
7000E GC/TQ with the HydroInert ion source. Separation
was performed on an Agilent J&W VF-WAXms GC, 60 m
× 0.25 mm, 0.25 µm capillary column (part number
CP9207). Alternatively, the same parameters can be
applied using a midcolumn backflush configuration with
two Agilent J&W VF-WAXms GC, 30 m × 0.25 mm, 0.25 µm
capillary columns (part number CP9205) at flows of
1 and 1.2 mL/min, respectively. The backflush setup was
also evaluated. Tables 1 and 2 provide the GC and MS
parameters, respectively.
The GC/TQ was operated in dynamic multiple reaction
monitoring (dMRM) mode. The MRM transitions for all nine
impurities were developed using the Agilent MassHunter
Optimizer for GC/TQ and used for data acquisition (Table 3).
3
Software and data integrity
Agilent MassHunter Workstation, including MassHunter
Acquisition 13.0 software for GC/MS and MassHunter
Quantitative Analysis 12.1 software, was used for data
acquisition and analysis. The OpenLab Electronic Content
Management (ECM) XT configuration provides capabilities
to facilitate compliance with various national and EU
electronic record regulations. The automated tools and
processes include the ability to create users with affiliated
permissions, generation of audit trails, and remote data
storage to minimize the risk of data breach or loss. OpenLab
ECM XT, employed with the MassHunter application, provided
a flexible data management solution with a single point of
access to data generated from instruments, data systems,
and laboratory software. With access to data from a storage
location, analysts can collaborate without compromising data
integrity and can create consistent processes for workflows.
Table 2. MS parameters.
Parameter Value
MS System Agilent 7010 Series GC/TQ with HES and
Agilent 7000E GC/TQ with HydroInert (HI) ion source
Mode Electron impact, 70 eV (on both HES and HI ion source)
Source Temperature 250 °C
Quadrupole Temperature Q1 and Q2 = 150 °C
MS1 and MS2 Resolution All compounds unit
Collision Gas Flow Nitrogen at 1.5 mL/min
Quench Gas Flow Helium at 2.25 mL/min when using helium carrier gas;
Switched off when using hydrogen carrier gas
Table 1. GC parameters.
Parameter Value
GC System Agilent 8890 GC system
MMI Injection Mode Pulsed splitless: 15 psi until 0.5 min
Inlet Temperature 250 °C
Inlet Liner Ultra Inert, splitless, single taper, glass wool (p/n 5190-2293)
Oven Temperature
Program
40 °C (1.5 min)
20 °C/min to 200 °C (0 min)
60 °C/min to 250 °C (3 min)
Total Run Time 13.33 min
MS Transfer Line
Temperature
250 °C
Injection Volume 2 µL
GC Column
VF-WAXms
Helium: 30 m × 0.25 mm, 0.25 µm column (p/n CP9205)
Hydrogen: 60 m × 0.25 mm, 0.25 µm column (p/n CP9207)
and midcolumn backflushing with two 30 m × 0.25 mm,
0.25 µm columns
Carrier Gas Hydrogen 1 mL/min (for HES and HydroInert) or
Helium 1.2 mL/min (for HES)
Compound
Retention Time
(min) MRM Transition CE
NDMA-D6 8.437 80 & 50 5
NDMA 8.448
74 & 44.1 6
74 & 42.1 24
43.1 & 42.1 10
NMEA 8.767
87.9 & 71 4
87.9 & 42.1 24
43.1 & 42.1 10
NDEA 8.969
101.9 & 85.1 4
101.9 & 56 20
101.9 & 44.1 14
NEIPA 9.198
115.9 & 99 6
115.9 & 44 16
71 & 56 6
NDIPA 9.366
130 & 88 6
130 & 71 16
130 & 42.1 12
NDPA 9.832
130 & 113.1 2
101 & 70 2
70 & 43.1 6
NDBA 10.796
158 & 141.1 4
158 & 99.1 10
116 & 99.1 4
84 & 56 22
NPIP 11.088
113.9 & 97.1 8
113.9 & 84.1 8
113.9 & 55 26
113.9 & 42.1 24
Table 3. Quantitative/qualitative transitions (dMRM-based).
Results and discussion
Spectral match quality
Full scan spectra for each of the eight analytes were acquired
using helium or hydrogen carrier gas, then compared against
the NIST library. Shown in Figure 1, excellent match scores
(> 90) were obtained using hydrogen carrier gas with the
7010 Series GC/TQ-HES system. Good match scores (> 80)
were obtained when using hydrogen carrier gas with the
7000E GC/TQ-HydroInert source setup. Higher average match
scores with the HES could be attributed to higher response
due to enhanced sensitivity.
Maintaining the integrity of mass spectra enabled the use of
identical MRM transitions for the helium and hydrogen carrier
gas methods (Table 3).
4
6.25 6.50 6.75 7.00 7.25 7.50 7.75 8.00 8.25 8.50 8.75 9.00 9.25 9.50 9.75 10.00 10.25 10.50 10.75 11.00 11.25 11.50 11.75 12.00 12.25
100%
100%
100%
Compound name = 1-Butanamine, N-butyl-N-nitrosoCAS# = 924-16-3
Compound name = 2-Propanamine, N-(1-methylethyl)-N-nitrosoCAS# = 601-77-4
Compound name = 1-Propanamine, N-nitroso-N-propyl -
CAS# = 621-64-7
Compound name = Piperidine, 1-nitrosoCAS# = 100-75-4
Compound name = Ethanamine, N-ethyl-N-nitrosoCAS# = 55-18-5
Compound name = Ethanamine, N-methyl-N-nitrosoCAS# = 10595-95-6
Compound name = N-Nitrosodimethylamine
CAS# = 62-75-9
= 92.5 Match factor
= 92.4 Match factor
= 84.2 Match factor
= 84.5 Match factor
= 81.8 Match factor
= 88.0 Match factor
Match factor = 94.1
NIST matching with He carrier gas using
an Agilent 7010 Series GC/TQ with HES
8.40 8.60 8.80 9.00 9.20 9.40 9.60 9.80 10.00 10.20 10.40 10.60 10.80 11.00 11.20 11.40 11.60 11.80 12.00 12.20 12.40 12.60 12.80 13.00 13.20 Compound name = 2-Propanamine, N-(1-methylethyl)-N-nitrosoCAS# = 601-77-4 Match factor = 96.1
Compound name = 1-Propanamine, N-nitroso-N-propylCAS# = 621-64-7
Match factor = 94.7
Compound name = 1-Butanamine, N-butyl-N-nitrosoCAS# = 924-16-3
Match factor = 91.7
Compound name = 2-Propanamine, N-ethyl-N-nitrosoCAS# = 16339-04-1
Match factor = 90.5
Compound name = Piperidine, 1-nitrosoCAS# = 100-75-4
Match factor = 92.2
Compound name = Ethanamine, N-ethyl-N-nitrosoCAS# = 55-18-5
Match factor = 97.4
Compound name = Ethanamine, N-methyl-N-nitrosoCAS# = 10595-95-6
Match factor = 95.7
Compound name = N-Nitrosodimethylamine
CAS# = 62-75-9
Match factor= 94.4
NIST matching with H2
as carrier gas using
an Agilent 7010 Series GC/TQ with HES
8.50 8.75 9.00 9.25 9.50 9.75 10.00 10.25 10.50 10.75 11.00 11.25 11.50 11.75 12.00 12.25 12.50 12.75 13.00 13.25 13.50 13.75 14.00 14.25 Compound name = 1-Butanamine, N-butyl-N-nitroso- CAS# = 924-16-3 Match factor = 87.1
Compound name = 2-Propanamine, N-(1-methylethyl)-N-nitrosoCAS# = 601-77-4
Match factor = 89.2
Compound name = 1-Propanamine, N-nitroso-N-propylCAS# = 621-64-7
Match factor = 87.8
Compound name = Piperidine, 1-nitrosoCAS# = 100-75-4
Match factor = 85.7
Compound name = Ethanamine, N-ethyl-N-nitrosoCAS# = 55-18-5
Match factor= 92.8
Compound name = N-Nitrosodimethylamine
CAS# = 62-75-9
Match factor = 96.2
NIST matching with H2
as carrier gas using
an Agilent 7000E GC/TQ with HydroInert source
A
B
C
Acquisition time (min)
Acquisition time (min)
Acquisition time (min)
Figure 1. Separation of eight nitrosamine impurities using helium carrier gas with the HES (A), hydrogen carrier gas with the HES (B), and hydrogen carrier gas
with the HydroInert source (C). The insets show the associated NIST match scores for the acquired full scan mass spectra.
5
Linearity
Method calibration performance for the eight impurities
was demonstrated over the range of 0.3 to 50 ng/mL with
R2
> 0.99 (Table 4). The lowest calibration level was the
concentration where the ion ratios for the qualifier ions
passed the ion ratio criteria. When using hydrogen carrier gas,
detection limits of 3 ng/mL or lower were achieved with both
the 7010 Series GC/TQ-HES and 7000E GC/TQ-HydroInert
source setups.
Limits of quantification
According to the latest regulatory directives6,7, the limit
of quantification (LOQ) must not exceed the established
acceptable limit for the specific nitrosamine impurity
measured. When a single analytical method is used to assess
various nitrosamines, the method’s selectivity at the LOQ
for each individual nitrosamine must be validated. The use
of methods with LOQs at or below 0.03 ppm is a common
pharmaceutical industry requirement for drug substances and
drug products.
Sample preparation involves a 10x dilution (500 mg extracted
with 5 mL of dichloromethane). Recovery studies that spiked
the drug substances at 30 ppb (0.03 ppm) resulted in extracts
with a concentration of 3 ng/mL for each nitrosamine
impurity measured in this study. This concentration
complied with the typical 30 ppb LOQ requirement for each
nitrosamine impurity.
With the 7010 Series GC/TQ, the carrier gas was switched
between helium and hydrogen over a span of six months
and the results were compared across three of those carrier
gas changes. Figure 2 illustrates that consistent spectral
fidelity, calibration response, ion ratios, and sensitivity were
observed. The spectral fidelity remained unaffected over
the measurement timeframe and generated consistently
high library match scores for all analytes (> 90). The
ability to maintain spectral fidelity and produce consistent
calibration response, ion ratios, and sensitivity after changing
the carrier gas establishes the method’s suitability for
practical applications.
Table 4. Calibration levels for the different carrier gases and ion sources.
Compound
Calibration Range (ng/mL) United States Pharmacopeia (USP) S/N of Quantifier Transition at Lowest Calibration Level
Agilent 7010 Series
GC/TQ with He
Carrier Gas
Agilent 7010 Series
GC/TQ with H2
Carrier Gas
Agilent 7000E
GC/TQ with
HydroInert Source
Agilent 7010
Series GC/TQ
with He Carrier Gas
Agilent 7010
Series GC/TQ
with H2
Carrier Gas
Agilent 7000E
GC/TQ with
HydroInert Source
NDMA 0.1 to 50 0.5 to 50 3 to 50 > 12 > 10 > 10
NMEA 0.2 to 50 0.3 to 50 1 to 50 > 200 > 12 > 40
NDEA 0.05 to 50 0.5 to 50 1 to 50 > 20 > 10 > 100
NEIPA 0.05 to 50 1 to 50 3 to 50 > 50 > 80 > 90
NDIPA 0.05 to 50 0.5 to 50 1 to 50 > 80 > 10 > 60
NDPA 0.1 to 50 0.3 to 50 3 to 50 > 100 > 10 > 10
NDBA 0.1 to 50 1 to 50 3 to 50 > 60 > 10 > 20
NPIP 0.1 to 50 1.25 to 50 3 to 50 > 60 > 10 > 10
6
NIST Matching 96.9
First set of results with
H2
carrier gas
NIST Matching 94.4
Second set of results with
H2
carrier gas after
swapping to He carrier gas
and back to H2
carrier gas
Figure 2. Consistent library match scores, calibration, and ion ratios were obtained using hydrogen carrier gas before and after using helium gas with an
Agilent 7010 Series GC/TQ.
7
Stability
The stability of the results for 150 consecutive injections was
previously examined using helium carrier gas (as described
in 5994-4618EN8
). In this work, the same assessment was
conducted with hydrogen carrier gas (Figure 3). The RSDs
(calculated with respect to absolute areas) were < 10% for
all the analytes, and the calculated concentration RSDs
(after internal standard correction) were < 7%. This indicated
long-term stability of response and applicability of the method
for routine analysis.
Figure 3. Peak area trend (using an Agilent 7000E GC/TQ with HydroInert source) for a nitrosamine impurity recovery sample at 30 ppb (with respect to the drug
substance). The plot was created using the metric plot feature in Agilent MassHunter Quantitative Analysis software.
9 15 21 27 33 39 45 51 57 63 69 75 81 87 93 99 105 111 117
-0.1
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
Area - N-Nitrosodimethylamine (NDMA)
Area - Ethanamine, N-methyl-N-nitroso- (NMEA)
Area - Ethanamine, N-ethyl-N-nitroso- (NDEA)
Area - 2-Propanamine, N-ethyl-N-nitroso- (NEIPA)
Area - 2-Propanamine, N-(1-methylethyl)-N-nitroso- (NDIPA)
Area - 1-Propanamine, N-nitroso-N-propyl- (NDPA)
Area - 1-Butanamine, N-butyl-N-nitroso- (NDBA)
Area - Piperidine, 1-nitroso- (NPIP)
×103
8
a Purged Ultimate Union (PUU) is installed between two
identical 30 m columns (0.25 mm × 0.25 µm). Consistent
results, including RTs and peak areas, were obtained when
using the midcolumn backflush configuration. Figure 4 shows
an example of the RTs and peak areas obtained after 25, 50,
and 100 consecutive sample runs.
Midcolumn backflush
With a dilution factor of 10, a considerably high amount of
drug substance matrix was introduced into the analytical
system. This can result in RT shifts and a gradual decrease
in peak response. To overcome this issue, the midcolumn
backflush capability can be used. In this configuration,
8.70 8.75 8.80 8.85 8.90 8.95 9.00
0
1
2
3
8.839 min
615.26
×102
Counts
After 25 sample runs
8.70 8.75 8.80 8.85 8.90 8.95 9.00
0
1
2
3 8.839 min
561.39
×102
Counts
After 50 sample runs
8.70 8.75 8.80 8.85 8.90 8.95 9.00
0
1
2
3 8.845 min
553.44
×102
Counts
After 75 sample runs
0
1
2
3
8.70 8.75 8.80 8.85 8.90 8.95 9.00
8.845 min
575.53
×102
Acquisition time (min)
Counts
After 100 sample runs
8.70 8.75 8.80 8.85 8.90 8.95 9.00
0
1
2
8.839 min
468.37
First injection of 0.03 ppm
recovery sample
×102
Counts
Figure 4. Consistent retention times (RTs) and peak areas were obtained after 25, 50, and 100 consecutive
sample runs when the mid-column backflush setup was used.
www.agilent.com
DE18669787
This information is subject to change without notice.
© Agilent Technologies, Inc. 2024
Printed in the USA, May 16, 2024
5994-7438EN
Recoveries
Sample recoveries were calculated by fortifying the drug
substances valsartan, irbesartan, losartan, and olmesartan
at 0.03 ppm. Recoveries were satisfactory and ranged from
80 to 120% when using hydrogen carrier gas on both the HES
and HydroInert source setups.
Conclusion
Using hydrogen carrier gas on an Agilent 8890 GC system
with either an Agilent 7010 Series GC/TQ (HES) or an Agilent
7000E GC/TQ (HydroInert source) demonstrated excellent
performance for the determination of eight nitrosamine
drug impurities in sartan drug products and substances.
Performance was validated at 0.03 ppm with acceptable
recovery and long-term repeatability. Both the 7000E GC/TQ
with the HydroInert source and the 7010 Series GC/TQ
with the HES source facilitated the ability of the system
to achieve the required detection limits. The integration
of OpenLab ECM XT with MassHunter Acquisition 13.0
streamlines data management, providing analysts with
centralized access to instrument-generated data, fostering
collaboration, maintaining data integrity, and optimizing
workflow processes.
References
1. Agilent EI GC/MS Instrument Helium to Hydrogen
Carrier Gas Conversion, Agilent Technologies user guide,
publication number 5994-2312EN, 2020.
2. Quimby, B. D.; Andrianova, A. A. Volatile Organic
Compounds Analysis in Drinking Water with Headspace
GC/MSD Using Hydrogen Carrier Gas and HydroInert
Source. Agilent Technologies application note, publication
number 5994-4963EN, 2022.
3. Quimby, B. D.; Haddad, S.; Andrianova, A. A. Analysis
of PAHs Using GC/MS with Hydrogen Carrier Gas and
the Agilent HydroInert Source. Agilent Technologies
application note, publication number 5994-5711EN, 2023.
4. Haddad, S.; Quimby, B. D.; Andrianova, A. A. GC/MS/MS
Analysis of PAHs with Hydrogen Carrier Gas Using the
Agilent HydroInert Source in a Challenging Soil Matrix.
Agilent Technologies application note, publication number
5994-5776EN, 2023.
5. Miles, L.; et al. EPA TO-15 Analysis Using Hydrogen
Carrier Gas and the Agilent HydroInert Source.
Agilent Technologies application note, publication number
59945359EN, 2022.
6. U.S Department of Health and Human Services. Food
and Drug Administration Center for Drug Evaluation and
Research. Control of Nitrosamine Impurities in Human
Drugs: Guidance for Industry. February 2021. https://
www.fda.gov/media/141720/download
7. European Medicines Agency. Questions and Answers
for Marketing Authorization Holders/Applicants on the
CHMP Opinion for the Article 5(3) of Regulation (EC) No
726/2004 Referral on Nitrosamine Impurities in Human
Medicinal Products. 15 January 2024. https://www.ema.
europa.eu/en/documents/referral/nitrosamines-emeah-a53-1490-questions-answers-marketing-authorisationholders/applicants-chmp-opinion-article-53-regulationec-no-726/2004-referral-nitrosamine-impurities-humanmedicinal-products_en.pdf
8. Dasgupta, S.; Dhyani, V; Churley, M. Quantification of Nine
Nitrosamine Impurities in Sartan Drugs Using an Agilent
GC-TQ. Agilent Technologies application note, publication
number 5994-4618EN, 2022.
Brought to you by
Download This App Note for FREE Now!
Information you provide will be shared with the sponsors for this content. Technology Networks or its sponsors may contact you to offer you content or products based on your interest in this topic. You may opt-out at any time.
Experiencing issues viewing the form? Click here to access an alternate version