Pharmaceutical manufacturers face significant challenges when analyzing low-volatility Class 2 residual solvents, as conventional static headspace gas chromatography methods struggle with these compounds.
While alternative techniques exist, none provide a comprehensive solution for detecting all Class 2 residual solvents. Molecular rotational resonance (MRR) spectroscopy offers a breakthrough approach by leveraging the unique rotational energy transitions of molecules in the microwave region to achieve extraordinary chemical selectivity without chromatographic separation.
This article explores how MRR spectroscopy enables direct analysis of complex mixtures, reducing analysis times and eliminating coelution concerns.
Download this article to learn:
- How MRR's chemical selectivity enables direct analysis of residual solvents
- Why MRR can detect low-volatility Class 2 solvents that conventional methods cannot
- How MMR supports continuous manufacturing and sampling across pharmaceutical development
STIMULI TO THE REVISION PROCESS
Stimuli articles do not necessarily reflect the policies of the USPC or the USP Council of Experts
Direct Analysis of Class 2 Residual Solvents Using Molecular Rotational Resonance Spectroscopy
Alexander V. Mikhonina
, Reilly E. Sonstroma
, Justin L. Neilla
, and Edmond Bibab
Correspondence should be addressed to: Edmond Biba, Senior Principal Scientist, US Pharmacopeia, 12601 Twinbrook Parkway, Rockville, MD
20852-1790; email: exb@usp.org.
ABSTRACT
Molecular rotational resonance (MRR) spectroscopy utilizes the microwave region of electromagnetic radiation, which, in terms of energy,
corresponds to transitions between pure rotational energy levels of molecules. Combined with adiabatic cooling in a gas phase, MRR
measurement provides a pure rotational spectrum of a molecule's ground vibrational state, which has absolute frequencies and extremely narrow
linewidths. Such a spectrum is unique for each chemical structure because it precisely reflects the absolute positions in space for all atoms in a
molecule and the corresponding three-dimensional mass distributions. Any small change in a position in space or the mass of any atom(s)
between two different substances will result in a well-resolved change in their MRR spectra. Hence, MRR offers unparalleled chemical selectivity.
This unique selectivity gives MRR a key advantage: mixtures, including mixtures of all types of isomers, can be analyzed directly without the
need for chromatographic separation. As such, MRR has the potential to reduce the time of analyses of mixtures containing diverse volatile
chemicals, including residual solvents listed in Residual Solvents 〈467〉.
This Stimuli article presents the development and validation of a continuous headspace-MRR method for analysis of selected Class 2 (moderate
toxicity) residual solvents (per 〈467〉), including all of the low-volatility solvents from USP Residual Solvents Class 2—Mixture C RS. Solvents for
this study were selected based on diversity of their boiling points to demonstrate MRR feasibility for analysis of a representative range of
solvents. The data presented suggest that MRR can meet ICH and USP requirements for analysis of most of Class 2 and Class 3 residual solvents,
and half of Class 1 residual solvents.
An attractive feature of using MRR for residual solvent analysis as an orthogonal technique is that it can bridge analytical gaps of conventional
methods for analysis of low-volatility solvents such as those from USP Residual Solvents Class 2—Mixture C RS, dimethyl sulfoxide (DMSO),
water-soluble acids, volatile amines, and others. Further, MRR is an online-capable technique that can be used for analysis of diverse volatile
molecules not only in laboratory settings but also on the process floor. Therefore, MRR can be used at any stage of pharmaceutical development
and manufacturing including support for process analytical technology (PAT) and quality by design (QbD) initiatives and continuous
manufacturing.
INTRODUCTION
Residual Solvents 〈467〉 is the USP documentary standard for the control of residual solvents in pharmaceutical products. It has information on
the categorization of residual solvents by risk assessment, limits of residual solvents in drug products and dietary supplement products, control
strategy, and analytical procedures for evaluating the levels of all Class 1 and most Class 2 residual solvents. These analytical procedures are
based on static headspace gas chromatography (SH-GC) with a flame ionization detector (FID) and were validated for all Class 1 solvents (with
severe toxicity) and most Class 2 solvents (with less severe toxicity).
The SH-GC methods in 〈467〉 cannot detect several residual Class 2 solvents, which are the components of USP Residual Solvents Class 2—
Mixture C RS (2-methoxyethanol; 2-ethoxyethanol; N,N-dimethylacetamide; N,N-dimethylformamide; ethylene glycol; formamide; Nmethylpyrrolidone; and sulfolane), as these solvents have lower vapor pressures than most of the other Class 1 and 2 solvents. As stated in
〈467〉, “those solvents should be determined using an alternative method that has been appropriately validated to demonstrate that the method
is suitable for its intended purpose”. Generally accepted guidances for validation of compendial and alternative procedures include Validation of
Compendial Procedures 〈1225〉 and ICH Q2(R2) (1). Residual Solvents—Verification of Compendial Procedures and Validation of Alternative
Procedures 〈1467〉 provides additional guidance on the specific validation criteria for alternative methods used for identification and quantification
of residual solvents.
The only type of gas chromatography (GC) method reported in the literature that is capable of analyzing all the solvents in USP Residual
Solvents Class 2—Mixture C RS utilizes direct injection GC (2–3). However, this method, despite its simplicity, is not a reliable option for analysis
of residual solvents. The biggest disadvantage of direct injection GC is that all formulation matrix components, including active pharmaceutical
ingredients and excipients, are also injected into the GC system. Nonvolatile or corrosive sample matrix components, if present, can cause
injector contamination, as well as column contamination and deterioration (4–6). Even if there are no nonvolatile or corrosive components in the
injected sample matrix, the matrix must be fully eluted prior to the next injection, resulting in an increased analysis time.
A few alternative methods, including reverse-phase HPLC (7), SPME-GC (8), SPME-GC-MS (9), and SIFT-MS (10), that are capable of analyzing
some of the solvents in USP Residual Solvents Class 2—Mixture C RS have also been reported. However, to our knowledge, none of these
alternative methods can provide a convenient and universal solution for analysis of all the solvents represented in USP Residual Solvents Class 2
—Mixture C RS.
Therefore, we studied the feasibility of using molecular rotational resonance (MRR) spectroscopy and validated its suitability as a convenient
universal solution for components of USP Residual Solvents Class 2—Mixture C RS, as well as several selected residual solvents from USP
Residual Solvents Class 2—Mixture A RS and USP Residual Solvents Class 2—Mixture B RS.
MRR is a spectroscopic technique that utilizes microwave radiation to probe the pure molecular rotation of compounds in gas phase, for both
pure compounds and mixtures of them. The compounds, both pure and in mixtures, can be precisely and unambiguously identified through their
rotational angular momentum transitions in the gas phase. Simplistically, the rotational angular momentum is a product of the principal moments
of inertia of a molecule and the angular velocities of rotation around its principal axes. Hence, the molecular rotational transitions are quantized
as determined by the principal moments of inertia that precisely reflect three-dimensional mass distributions of molecules. Even in the most
complex general case of asymmetric top molecules, where all three principal moments of inertia are not equal to zero, the process of solving the
molecular Hamiltonian using modern quantum chemistry methods is well-developed and highly accurate (11). Furthermore, as the principal
moments of inertia can be directly and accurately calculated for any given equilibrium molecular geometry, it is straightforward to theoretically
predict and assign the corresponding molecular MRR spectral patterns.
These MRR spectral patterns are information-rich, well-defined, reproducible, highly characteristic, and extremely sharp with lines routinely
determined to a relative frequency precision of >106
(ratio of frequency to error in frequency, e.g., <10 kHz frequency error on a line frequency
of 10 GHz or greater). Even the molecules with highly similar structures, such as regioisomers (12), diastereomers (13), and isotopomers (14–
15), have distinct principal moments of inertia to produce unique MRR fingerprints, with essentially no spectral overlaps. Enantiomers, which
have identical moments of inertia, can still be easily discriminated using gaseous chiral tag molecules that convert the original enantiomers into
the MRR-distinguishable weakly-bonded diastereomeric complexes directly in a gas phase (16–17). This excellent selectivity means that virtually
all compounds, including all types of isomers, can be resolved by MRR without chromatographic separation.
In recent years, advances in measurement technology have enabled the characterization of increasingly complex systems including crude
reaction mixtures (13,20) pharmaceutical raw material impurities (12), isotopically labeled mixtures (15,18–19), and mixtures of enantiomers
(19). MRR has also been extended to online measurement applications (13,21). Its primary limitations are that analytes must have a permanent
dipole moment and that the analyte must be analyzed in the gas phase (11).
This article demonstrates that MRR with continuous headspace sampling is capable of not only accelerating the analysis of residual solvents via
direct analysis but also bridging the analytical gap of SH-GC for analysis of low-volatile residual solvents from USP Residual Solvents Class 2—
Mixture C RS. MRR meets the analytical detection requirements for a wide array of volatile and semi-volatile impurities for complying with the
acceptance criteria using a single instrumental setup. The continuous headspace sampling protocol involves flowing neon carrier gas over a
heated sample, from which vapor phase components are introduced into the gas stream. The resulting mixture enters a vacuum chamber
through a pulsed supersonic valve, after which the MRR spectra are recorded. Each of the MRR-detectable USP residual solvents, including each
of those in USP Residual Solvents Class 2—Mixture C RS, has a distinct MRR spectrum that can be used for extraordinarily selective identification
as well as quantitation in the drug product. Each solvent in the sample can be analyzed with a variable measurement time depending on the
measurement sensitivity required, and multiple solvents can be analyzed in combination without the risk of misidentification of any of them. The
method can also be readily extended to include more volatile residual solvents in any of the other classes as well as other residual impurities.
The use of alternative procedures for residual solvent analysis, even though generally permissible, requires proper validation to demonstrate
the adequacy of these procedures and the suitability of their analytical metrics for the purpose. More details about validation procedures,
parameters, and definitions that are widely accepted in the pharmaceutical industry can be found in 〈1467〉 and ICH Q2(R2) guidance (1). Briefly,
for quantitative methods it is generally required to demonstrate the method selectivity, linearity and range, repeatability, limit of quantitation
(LOQ), and accuracy and recovery.
The MRR prototype used in this study was capable of meeting USP and ICH selectivity, sensitivity, linearity, and range requirements for all
solvents studied, including solvents poorly suited for GC analysis such as thosein theUSP Residual Solvents Class 2—Mixture C RS. Analysis
repeatability and accuracy/recovery requirements were met for most of the studied residual solvents as well. With upcoming technology
improvements, MRR is expected to fully comply with USP and ICH analytical and method validation requirements for most of residual solvents
from Classes 1, 2, and 3.
In addition to bridging a gap of SH-GC for analysis of the solvents represented in USP Residual Solvents Class 2—Mixture C RS, MRR is also
capable of direct analysis of other residual solvents that are poorly suited for GC analysis, such as acetic acid, formic acids, and dimethyl
sulfoxide (DMSO). Other challenging volatile impurities that are not included in〈467〉 can also be analyzed using the presented headspace-MRR
method without any derivatization required. Examples include but are not limited to ethylene oxide, formaldehyde, acetaldehyde, carboxylic
acids, volatile amines, and sulfur-containing species.
INSTRUMENTAL METHODS
Continuous Headspace Method
MRR uses continuous headspace sample introduction that allows for effective sampling of large headspace volumes, which improves the
sensitivity for analytes that are present at very low concentrations. Furthermore, the measurement time can be tailored for each analyte in a
sample to meet the sensitivity requirements for that analyte and to optimize the measurement cycle time in case of multi-analyte analyses.
Figure 1 shows a diagram of a continuous headspace-sampling MRR spectrometer, including both the sampling and measurement parts.
Research-grade neon (>99.999% purity) is used as the carrier gas, which flows over a sealed headspace vial where the sample (diluted, if
necessary, in a suitable diluent) has been allowed to equilibrate at a specific temperature. This temperature is chosen to optimize sensitivity for
the analyte(s) to be characterized. The carrier gas inlet pressure is held to a typical constant value of +5 psig. The resulting mixture of neon and
volatilized analytes flows through a heated transfer line and is introduced into a high vacuum chamber (<10−5
Torr) through a pulsed solenoid
valve. The vacuum pump (a turbomolecular pump with 950 L/s pumping speed, backed by a dry-compression backing pump) maintains a
chamber pressure in the ~10−4
Torr pressure range during measurements, which is suitable for sensitive MRR measurements.
The role of the pulsed solenoid valve is to create a supersonic expansion due to the rapid speed with which it opens and the pressure difference
between the vacuum and the analyte mixture being injected. These pulsed supersonic valves have been in regular use in rotational spectroscopy
instruments for over 40 years (22). This allows adiabatic cooling of the rotational and vibrational states of the molecule, which greatly increases
the MRR signal strength in the microwave region of the spectrum for the vast majority of the analytes in this study. These valves also prepare
isolated molecules capable of free rotation, so that the transition frequencies and line shapes are independent of any other components in the
sample, allowing for excellent selectivity and quantitation.
Click image to enlarge
Figure 1. Schematic diagram of the continuous headspace-sampling MRR instrument. See the text for further description.
Threshold Total Vapor Pressure for MRR Measurements
For each component of the sample mixture in the sealed headspace vial, including the diluent, the equilibrium headspace concentration of that
component in the analyzed gas mixture is directly proportional to its partial vapor pressure above the solution. We have found that for
quantitative MRR measurements, it is important to keep the total gas-phase vapor pressure below about 1 Torr (matrix-dependent), which
corresponds to total concentrations of all volatile components of approximately 0.1% in the gas stream. At higher analyte concentrations, MRR
signals exhibit a nonlinear relationship with concentration due to the formation of noncovalent complexes in the supersonic expansion;
noncovalent complexes exhibit rotation as a unit (rather than as free molecules) and so do not contribute to the total signal of that particular
analyte.
It should be noted that about 1 Torr value is given only as a general value. In highly polar sample matrixes, where probability of forming the
noncovalent complexes in a gasphase is high, the actual total vapor pressure threshold value could be less than 1 Torr. In contrast, in relatively
nonpolar sample matrices, this vapor pressure threshold value may be greater than 1 Torr.
Click image to enlarge
Figure 2. Example of experimental determination of the maximum allowable headspace temperature for MRR analysis. On the left, a plot of the
observed DMSO signals from a 20-mg/mL solution in sulfolane against the estimated DMSO partial pressure is given. This partial pressure is
calculated assuming Raoult’s Law (see Appendix). On the right, the vapor pressure contributions to the solution as a function of temperature are
given.
To illustrate the importance of the total vapor pressure of all components for headspace-MRR analysis, Figure 2 shows MRR signals of 20 µg/mL
DMSO in sulfolane (>99% purity), as well as the total vapor pressure of the system (predominately, sulfolane), at different headspace
temperatures. As evident, at headspace temperatures below about 80°, MRR signals are directly proportional to the partial vapor pressure of
DMSO. At headspace temperatures greater than about 80°, there is a deviation from the linear response attributed to the formation of
noncovalent complexes of DMSO (presumably, with sulfolane). As such, 80° is the maximum headspace temperature for this particular MRR
analysis. This temperature is chosen because, first, the MRR signal is verified to have a linear response to trace-level DMSO concentrations of up
to 20 µg/mL in sulfolane solution, and, second, MRR sensitivity to solution-phase DMSO is greater at this temperature than at lower headspace
temperatures. As evident from the right panel of Figure 2, the total vapor pressure threshold value for DMSO in the sulfolane sample is about 0.4
Torr. A practical approach for how to estimate the optimal headspace temperature for quantitation of volatiles in a given sample matrix will be
discussed in Targeted MRR Method Development.
One consequence of the solution vapor pressure limitation is that it is typically preferable to choose a diluent that has low vapor pressure to
maximize the sensitivity of the headspace-MRR measurement. For this reason, in this work, we used an ionic liquid, 1-butyl-3-methylimidazolium
tetrafluoroborate, or BMIMBF , as a diluent, which has negligible vapor pressure.
MRR Measurement
The targeted MRR spectrometer used for the measurements in this study has been described in greater detail elsewhere (11). Briefly, a FabryPerot resonator consists of two spherical mirrors that are separated by approximately 30 cm. One of the mirrors is fixed and has the solenoid
valve mounted into it, as well as two feed-throughs for coupling microwave radiation in and out of the cavity. The other mirror is mounted on a
translation stage capable of precisely tuning the length of the cavity to a value specified by the instrument software. To perform a measurement
at a specified frequency, the movable mirror first tunes to a length where a stable cavity mode can be found. A Schottky diode detector is used to
confirm that the mirror meets the resonance condition, and small corrections can be made if necessary. Once this is confirmed, the pulsed
solenoid valve begins pulsing gas into the chamber at a typical repetition rate of 10 Hz, with each gas pulse having a typical duration of 1 ms.
Following each gas pulse, a series of microwave pulses are transmitted into the chamber that are coincident with the molecular line frequency
to be measured. Each pulse, with a typical duration of 1–4 µs, induces a macroscopic coherence between a portion of the molecules of that
analyte in the sample chamber. Once the pulse is switched off, the molecules emit a free induction decay signal at their characteristic MRR
transition frequency. This emission is downconverted, recorded on a digitizer, and Fourier transformed to yield the resulting frequency spectrum.
This measurement scheme is analogous to Fourier transform nuclear magnetic resonance (FT-NMR) spectroscopy. The length of a single
measurement can be configured by the user, with typical measurement lengths ranging from a few seconds to 10 min per analyte. Once the
measurement at a single frequency is complete, the cavity can be automatically retuned to a new frequency within a few seconds, and a new
analyte can be characterized. The power of the microwave pulses is automatically adjusted by software for each compound, as the optimal pulse
power depends on the dipole moment of the transition being measured.
For the validation study described in this Stimuli article, a database of residual solvents and MRR transition frequencies has been compiled.
Many of the residual solvents have had their rotational spectra characterized in the past by different laboratories, resulting in a set of transition
frequencies that could be used in this study without further modification (the literature references to these data are cited in Table 2). For solvents
that had not been previously characterized by MRR, their spectra were measured using a broadband chirped pulse Fourier transform microwave
(CP-FTMW) spectrometer (23). This spectrometer uses broadband horn antennas to transmit excitation pulses through the chamber to interact
with the molecular sample, which reduces the sensitivity of the measurement but enables much larger bandwidth measurements for new
compounds.
MRR Spectral Assignment
One of the most attractive features of MRR is that, as the spectrum is precisely determined by the three-dimensional structure of the molecule
in the gas phase, the molecular identity can be confirmed unambiguously by comparing it to theoretical calculations. Furthermore, the
unambiguous spectral assignments can be done directly in crude mixtures without requiring a reference sample, as soon as a list of candidate
structures matches the reality (13). In recent years, developments in quantum chemistry methods, in particular dispersion-corrected density
functional theory (24), have enabled accurate prediction of molecular equilibrium geometry, from which MRR spectroscopy parameters can be
accurately derived.
Figure 3 provides an example of an MRR spectrum assignment for tetralin, a Class 2 residual solvent that had not been previously characterized
with MRR. In this figure, the black positive-going trace in each panel represents the experimentally measured MRR spectrum. In the left sub-plot,
the red negative-going spectrum is simulated using the rotational constants determined from quantum chemical calculations. While there is not
an exact match between the line positions, the strong lines show a clear comparable pattern. The theoretical rotational constants can be refined
to yield a new simulation (the blue spectrum in the right panel) where the frequencies of the experimental and simulated spectra agree within
experimental precision (<10 kHz). The structure is verified by comparing the experimental and theoretically calculated rotational constants
(Table 1). The constants agree to within less than 0.3% so that the identity of the measured compound as tetralin can be unambiguously
confirmed.
Table 1. Comparison of Theoretical and Experimental Rotational Constants of Tetralina
Rotational Constants Experimental Fit Theory Error
A (MHz) 2776.032 2783.93 −0.28%
B (MHz) 1127.1022 1125.935 +0.10%
C (MHz) 829.8582 829.005 +0.10%
a Rotational constants were calculated using B3LYP GD3BJ/6-311++G(d,p) basis set.
4
Click image to enlarge
Figure 3. Assignment of tetralin’s MRR spectrum (black: measured MRR spectrum of tetralin; red: simulated MRR spectrum of tetralin using
theoretically calculated rotational constants; blue: simulated spectrum of tetralin using experimentally refined rotational constants).
Selectivity of MRR Measurements
In a targeted MRR spectrometer, the typical full-width at half-maximum (FWHM) of an individual transition is ~0.1–0.2 MHz, while the
frequency range of the instrument is 5000–18500 MHz, leading to approximately 105
measurement channels. A single molecule, depending on its
size and geometry, may have anywhere from a few to several hundred possible detection frequencies that are extremely narrow. As a result, it is
possible to identify overlap-free detection frequencies for individual species, even in complex multi-component mixtures.
The main plot of Figure 4 shows MRR spectra for five solvents included inUSP Residual Solvents Class 2—Mixture C RS, simulated with
experimental linewidth, along with the spectrum of DMSO, the diluent in this standard. The linewidth of the transition is not visible on this scale.
The insets show experimental measurements using the targeted spectrometer for each of the six components from this standard, showing that
each line is clear and unblended, and the chance of overlap is extremely low; in addition, if there is an unlikely event of a coincidental overlap,
each compound has multiple lines that could be used for the analysis. With this selectivity of MRR, the direct quantitative analysis of residual
solvent mixtures is straightforward.
Click image to enlarge
Figure 4. MRR selectivity for residual solvents from USP Residual Solvents Class 2—Mixture C RS. Main plot (bottom): simulated MRR spectra
for five solvents from USP Residual Solvents Class 2—Mixture C RS and DMSO diluent. The spectra are simulated using the experimental fits (see
MRR Spectral Assignment for more details). Top subplots: targeted MRR spectra of the same analytes. Note that ethylene glycol has a second line
of different species in the measurement window, but this does not interfere with ethylene glycol identification or quantification.
MATERIALS
The list of residual solvents used in this study is presented in Table 2. Out of the 33 Class 2 residual solvents listed in 〈467〉, five were not
included in the MRR library because they have no dipole moment: cyclohexane;trans-1,2-dichloroethene; 1,4-dioxane;p-xylene; and n-hexane. It
should also be noted that 24 out of 26 Class 3 residual solvents are also included in the targeted MRR library.
Table 2. List of Class 2 Residual Solvents (per 〈467〉) Included in the MRR Library
Solvent Name
Puritya
(%)
Boiling Point
(°C)
Respective USP Reference
Standard MRR Source
Frequency
(MHz)
2-Methoxyethanol 99.5 124
USP Residual Solvents Class 2
—Mixture C RS (25) 15609.95
2-Ethoxyethanol 99.5 135.6
USP Residual Solvents Class 2
—Mixture C RS (26) 15470.76
Dimethylacetamide
(DMAC) 99.8 165
USP Residual Solvents Class 2
—Mixture C RS (27) 14729.29
Dimethylformamide
(DMF) 99.9 153
USP Residual Solvents Class 2
—Mixture C RS (28) 14123.51
Ethylene glycol 99.5 197.3
USP Residual Solvents Class 2
—Mixture C RS (29) 10747.53
N-Methylpyrrolidone
(NMP) 99 202
USP Residual Solvents Class 2
—Mixture C RS
This Stimuli
article 15696.07
Formamide 99.5 210
USP Residual Solvents Class 2
—Mixture C RS (30) 9237.03
Sulfolane 99 285
USP Residual Solvents Class 2
—Mixture C RS (31) 15698.79
Acetonitrile 99.8 82
USP Residual Solvents Class 2
—Mixture A RS (32) 18398.00
Chlorobenzene 99.5 132
USP Residual Solvents Class 2
—Mixture A RS (33) 8350.01
cis-1,2-Dichloroethene 99.2 60
USP Residual Solvents Class 2
—Mixture A RS (34) 13604.4
Ethylbenzene 99.5 136
USP Residual Solvents Class 2
—Mixture A RS (35) 13144.75
Isopropylbenzene 99.5 152.4
USP Residual Solvents Class 2
—Mixture A RS (36) 8276.39
Methanol 99.8 64.7
USP Residual Solvents Class 2
—Mixture A RS (37) 12178.59
Methylcyclohexane 99.5 100.9
USP Residual Solvents Class 2
—Mixture A RS
This Stimuli
article 10366.27
Methylisobutyl ketone 99.3 116
USP Residual Solvents Class 2
—Mixture A RS (38) 14475.59
Dichloromethane 99.8 39.6
USP Residual Solvents Class 2
—Mixture A RS (39) 15906.1
Tetrahydrofuran 99.9 66
USP Residual Solvents Class 2
—Mixture A RS (40) 13385.79
Toluene 99.8 110.6
USP Residual Solvents Class 2
—Mixture A RS (41) 15834.19
m-Xylene 99.5 139
USP Residual Solvents Class 2
—Mixture A RS (42) 14708.15
o-Xylene 99.2 144
USP Residual Solvents Class 2
—Mixture A RS (43) 9214.11
Chloroform 99.5 61.2
USP Residual Solvents Class 2
—Mixture B RS (44) 13204.28
1,2-Dimethoxyethane 99.5 85
USP Residual Solvents Class 2
—Mixture B RS (45) 14691.54
2-Hexanone 99.5 128
USP Residual Solvents Class 2
—Mixture B RS (46) 15989.57
Nitromethane 99 100
USP Residual Solvents Class 2
—Mixture B RS (47–48) 16420.10
Solvent Name
Puritya
(%)
Boiling Point
(°C)
Respective USP Reference
Standard MRR Source
Frequency
(MHz)
Pyridine 99.5 115
USP Residual Solvents Class 2
—Mixture B RS (49–50) 14903.01
Tetralin 98.4 208
USP Residual Solvents Class 2
—Mixture B RS
This Stimuli
article 10817.28
Trichloroethene 99.5 87.2
USP Residual Solvents Class 2
—Mixture B RS (51) 13424.9
a From the certificates of analysis provided by suppliers.
Samples and Standards Preparation
Standard stock solution: 200 μL of USP Residual Solvents Class 2—Mixture C RS were added in a 5-mL volumetric flask and diluted with
BMIMBF to volume.
Sample solution: 500.0 mg of respective drug substance were accurately weighed and transferred into a 25-mL volumetric flask and diluted
with the ionic liquid to volume.
Spiked sample stock solutions: 200 μL of USP Residual Solvents Class 2—Mixture C RS were added in a 5-mL volumetric flask and diluted
with Sample solution (corresponding to 1000% of the USP limit for a 20-mg/mL solution of drug substance) to volume. Three sets of Spiked
sample stock solutions were prepared: acetaminophen, dextromethorphan, and diphenhydramine, respectively.
Standard solutions: Add the respective volume of Standard stock solution (see Table 3) into a 2-mL volumetric flask and dilute with the ionic
liquid to volume.
Spiked sample standard solutions: Add the respective volume of Spiked sample stock solutions (see Table 3) into a 2-mL volumetric flask
and dilute with Sample solution to volume.
The Standard solutions and Spiked sample standard solutions were prepared at five different concentration levels (see Table 3) corresponding to
50%, 75%, 100%, 150%, and 250% of the USP limit for a 20-mg/mL solution of drug substance.
Table 3. Preparation of Standard Solutions at Five Concentration Levels
Standard Solution ID Level 1 Level 2 Level 3 Level 4 Level 5
Percentage of USP Limit in Solution (%)a 50 75 100 150 250
Volume of Standard Stock Solution (µL) 100 150 200 300 500
a USP limit in solution is calculated assuming 20-mg/mL concentration of a drug substance.
As for the remaining solvent members of USP Residual Solvents Class 2—Mixture C RS (formamide and sulfolane), as well as the selected
solvent members of USP Residual Solvents Class 2—Mixture A RS and USP Residual Solvents Class 2—Mixture B RS, we prepared custom
standards as described in Tables 4 and Table 5. The Table 4 residual solvents were added into a 5-mL volumetric flask prefilled with 3 mL of
diluent and then diluted with a corresponding diluent to volume to prepare the stock solutions 1-A through 1-D. The Table 5 Standard stock
solution and Standard solutions were prepared sequentially by adding a specified volume of a specified stock solution into a 2-mL volumetric
flask prefilled with 0.5 mL of diluent and then diluting it with a relevant diluent to volume. As diluents, we used either the ionic liquid (BMIMBF )
for linearity studies or a solution of a pharmaceutical substance in the ionic liquid at a specified concentration for accuracy and recovery studies.
Solvents in both the Standard stock solution and Standard solutions were grouped by their boiling points to optimize MRR sensitivity.
Table 4. Preparation of Custom Stock Solutions 1-A through 1-Da
Stock
IDb
Residual
Solvent
Boiling
Point
(°C)
USP
Limit
(µg/g)
Substance
Concentration
(mg/mL)
Reference Standard
Concentration in Solution
Corresponding to the USP
Limit
Volume in
Stock
Solution 1
µg/mL ppmv (µL)
Stock
solution
1-A
cis-1,2-
Dichloroethene 60 1870 10 18.7 14.6 290
Methanol 64.7 3000 10 30 37.9 760
Tetrahydrofuran 66 720 10 7.2 8.1 160
Acetonitrile 82 410 10 4.1 5.2 105
Stock
solution
1-B
Nitromethane 100 50 10 0.5 0.44 9
Toluene 110.6 890 10 8.9 10.3 205
Pyridine 115 200 10 2 2.04 40
Stock
solution
Chlorobenzene 132 360 10 3.6 3.2 65
Ethylbenzene 136 368 10 3.68 4.3 85
4
4
Stock
IDb
Residual
Solvent
Boiling
Point
(°C)
USP
Limit
(µg/g)
Substance
Concentration
(mg/mL)
Reference Standard
Concentration in Solution
Corresponding to the USP
Limit
Volume in
Stock
Solution 1
µg/mL ppmv (µL)
1-C m-Xylene 139 1302 10 13.02 15.1 305
Stock
solution
1-D
Formamide 210 220 40 8.8 7.8 155
Sulfolane 285 160 40 6.4 5.1 100
a These solutions contain 2–4 residual solvents at concentrations corresponding to 4000× the USP limits. (See 〈467〉. All standards are diluted to a 5-mL volume. Solvents are
grouped by boiling points.)
b
Ionic liquid (BMIMBF ) or 10-to 40-mg/mL substance solution in the ionic liquid.
Table 5. Preparation of Custom Standard Solutions at Five Concentration Levels for Selected Solvents Listed in USP Residual
Solvents Class 2—Mixture A RS, USP Residual Solvents Class 2—Mixture B RS, and USP Residual Solvents Class 2—Mixture C RS as
Specified in Table 4and the Text
Solution ID
Stock
Solution 2
Stock
Solution 3
Level
1-A
Level
2
Level
3
Level
4 Level 5a
Concentration of Reference Standard in
Solution (% of USP Limit) 20,000 1000 20 25 50 100 200 400 500
Volume of Stock 1 (µL) 100 — — — — — — — —
Volume of Stock 2 (µL) — 100 — — — — — — —
Volume of Stock 3 (µL) — — 40 50 100 200 400 800 1000
a
Ionic liquid (BMIMBF ) or 10- to 40-mg/mL substance solution in the ionic liquid.
HEADSPACE-MRR METHOD DEVELOPMENT
Diluent Considerations
As discussed previously, the diluent for headspace-MRR analysis must meet two criteria. First, this diluent should be capable of dissolving a
drug substance of interest to the required concentration level (typically, within the 10-to50-mg/mL range). Second, the diluent should have less
than approximately 1 Torr of vapor pressure in the intended range of headspace temperatures. For this reason, an ionic liquid (1-butyl-3-
imidazolium tetrafluoroborate or BMIMBF ) was used in this work.
Solubility Studies
Prior to any MRR measurements, it was verified that all the pharmaceutical drug substances involved in this study, such as acetaminophen,
aspirin, ibuprofen, dextromethorphan, and diphenhydramine, can be fully dissolved in the BMIMBF ionic liquid diluent to the required levels that
vary during our study between 10 and 40 mg/mL. In some cases, sonication was required for dissolution.
In case of encountering solubility issues while using BMIMBF to dissolve a substance of interest, there are other commercial ionic liquid
options, including EMIM DEP (1-ethyl-3-methylimidazolium diethyl phosphate) that are reported in the literature to be capable of dissolving even
cellulose (52). Alternatively, small quantities of a low-volatile cosolvent, such as sulfolane, can be added to BMIMBF or other suitable ionic liquid
to resolve possible solubility issues.
Determining Optimal Headspace Temperature for MRR Analysis
For each Standard solution, the maximum temperature range was estimated by calculating the total vapor pressure of the solution following a
procedure described in the Appendix. All components of the solution were considered, including, for example, the DMSO used as the diluent in
the manufacture ofUSP Residual Solvents Class 2—Mixture C RS. We found that for this sample, DMSO had the largest contribution to the vapor
pressure of the solution. For the other stock solutions, which were prepared from individual solvent standards, the vapor pressure was limited by
the analytes themselves (since the ionic liquid diluent contributes negligible vapor pressure).
MRR Library for Residual Solvent Analysis
For each Class 2 solvent, a strong spectral line was chosen based on its signal intensity within the range of the MRR spectrometer used in this
study (6–18.5 GHz). These frequency values are given in Table 2. Almost all of these solvents have additional transitions in the MRR library that
are also suitable for targeted analysis. The library contains the full characterization of the MRR spectrum of each solvent in this frequency range,
compiled from the sources given in Table 2. The Class 2 solvents that were not previously in the MRR library were added through the procedure
described previously.
Even though spectral overlaps between different molecules are extremely rare for MRR, the measured or simulated spectra of all components of
interest can be overlayed on one plot to verify that the selected (strongest) lines are overlap-free in the mixture of interest, similar to that of five
solvents from USP Residual Solvents Class 2—Mixture C RS in Figure 4.
After targeted lines for each analyte of interest were selected, the MRR instrument ran an automated routine to optimize MRR parameters for
the selected line to maximize analysis sensitivity. To save time, this routine can be run using a sample mixture with the highest quantifiable
analyte concentrations and the optimal headspace temperature for the intended MRR method as discussed.
4
4
4
4
4
4
VALIDATION STUDY DESIGN
There are two main goals of this validation study. The first goal is to verify that headspace-MRR can bridge the analytical gap of conventional
SH-GC methods for analysis of low-volatile residual solvents from USP Residual Solvents Class 2—Mixture C RS. Furthermore, we evaluated the
possibility of MRR integration into existing residual solvent analysis workflows by preparing residual solvent standards using USP Residual
Solvents Class 2—Mixture C RS.
The second goal is to demonstrate MRR feasibility for analysis of most of Class 2 residual solvents. For this purpose, we selected several
diverse residual solvents from USP Residual Solvents Class 2—Mixture A RS, USP Residual Solvents Class 2—Mixture B RS, and USP Residual
Solvents Class 2—Mixture C RS that cover most of the boiling point range for Class 2 residual solvents. The selected solvents were grouped by
their boiling points and can be classified as high-volatility (1,2-cis-dichloroethene, methanol, tetrahydrofuran, and acetonitrile), medium-volatility
(nitromethane, toluene, pyridine), medium-to-low volatility (chlorobenzene, ethylbenzene, m-xylene), and very low volatility (formamide and
sulfolane).
It should be noted that for the second half of the feasibility study we utilized custom standards that were prepared using chemically pure
solvents, grouped by their boiling points, and ionic liquid as a diluent for pharmaceutical substances under study. The reason for using the
custom standards is that it gives an option to increase the headspace temperature (while still meeting the maximum allowable vapor pressure
requirement) to further reduce the time required to reach the desired measurement sensitivity.
The validation study described in this article was designed in accordance with the requirements specified in the ICH Q2(R2) guideline and
recommendations specified in 〈1467〉 for validation of nonchromatographic alternative procedures for residual solvents.
VALIDATION RESULTS
System Suitability
MRR is an extraordinarily selective analytical technique; the MRR spectral pattern of a molecule cannot be matched by the spectral pattern of
any other molecule. There are essentially no spectral overlaps between different residual solvent species or other constituents of the drug
substance matrix. For these reasons, the main factor that determines MRR suitability for the task is the analysis sensitivity that, in turn, can be
accurately predicted theoretically (11). As a result, once MRR sensitivity for analysis of a particular residual solvent is verified, no specific system
suitability test is required besides the routine MRR instrument operational qualification tests.
Specificity and Selectivity
ICH Q2(R2) guidance states that "the specificity or selectivity of an analytical procedure can be demonstrated through the absence of
interference, comparison of results to an orthogonal procedure or may be inherently given by the underlying scientific principles of the analytical
procedure". MRR is an inherently highly selective analytical technique, and spectral interferences between different species are very rare.
However, we recommend a few practical steps to further augment the analysis specificity.
First, if all volatile components of a sample matrix of interest are in the MRR library (including the target analytes, drug substance, impurities,
degradation products, excipients, and diluents), then overlaying their library spectra (including all detectable conformers) on the same plot can
be used to check for potential spectral overlaps.
Second, if some of the volatile matrix components or their MRR spectra are unknown or if spectral overlap is suspected, we recommend testing
those components by the MRR method using blank and spiked samples. If spectral overlap is confirmed for a transition, then another transition
from the MRR library can be used for the targeted analysis of this analyte(s).
Alternatively, it is possible to develop MRR methods that use multiple targeted lines for each analyte of interest. However, this approach will
result in increased analysis cycle time.
linearity and range
According to ICH Q2(R2), a minimum of five concentrations appropriately distributed across the range is recommended for assessing the
linearity. For residual solvents testing, the reportable range to be demonstrated is 50%–120% of the target concentration (i.e., 50%–120% of
the USP concentration limit for each residual solvent). Chapter 〈1467〉 recommends for nonchromatographic procedures that the coefficient of
determination of the linearity curve, R
2
, is not less than 0.90.
In the first part of our validation study, where USP Residual Solvents Class 2—Mixture C RS was used for sample preparation Table 3, the
demonstrated linear range is 50%–250% of the USP limit (Figure 5 and Table 6). In the second part of our study, where custom-made residual
solvent standards were used, the demonstrated linear range is at least 25%–400% of the USP limit (Figure 6, Figure 7, and Table 8). Pearson's
linear regression coefficient, R
2
, was equal to or greater than about 0.99 for most of the residual solvents and greater than about 0.97 for all
solvents. Therefore, the requirements for linearity and range were met for all listed residual solvents.
limit of quantitation
LOQs were estimated for each solvent from a corresponding linear slope and a standard deviation (SD) of six measurements of blank MRR
signals (see Table 6 and Table 8). The 600 s duration of blank measurements was chosen because 600 s represents the maximum practical
measurement time per analyte for headspace-MRR analysis. The mathematical formula for LOQ is provided below.
LOQ = 10 × SD /Slope
The ICH Q3C(R2) and 〈467〉 provide the concentration limits for residual solvents based on the permissible daily exposure (PDE) when the
maximum daily dose of the drug product does not exceed 10 g/day. The LOQs for all analytes (residual solvents) of this study were no more
than 50% of USP limits per 〈467〉 (see Table 6 and Table 8). Since the spectrometer used to perform the analysis measures selected rotational
transitions (corresponding to each analyte) separately, these quantitation limits can be easily and independently adjusted as needed in any
direction by shortening or increasing the measurement time for a specific analyte, as needed.
Repeatability
As stated in ICH Q2(R2), repeatability can be determined from either six measurements at 100% of target concentration or at least nine
measurements within a reportable range for the procedure (e.g., three concentrations and triplicate analysis). The recommended acceptance
Blank
criterion for validation of nonchromatographic alternative procedures in 〈1467〉 is a relative standard deviation of not more than 20% for each
solvent present.
For the first part of our study, we used three replicates at each of the 75%, 100%, and 150% (of the USP limit) concentration levels to
calculate the repeatability (see Table 6). For the second part of our study, we used three replicas at each of the 50%, 100%, and 200% (of the
USP limit) concentration levels (see Table 8). As evident from Table 6 and Table 8, the repeatability for most of residual solvents is less than
20% and meets the ICH requirements. The exceptions are dimethylacetamide (DMAC), tetrahydrofuran (THF), ethylbenzene, and m-xylene.
Accuracy
To evaluate accuracy and recovery, three spiked sample solutions were prepared for each concentration level for the three drug substances. For
the first part of this study, we used 20-mg/mL solutions of acetaminophen, dextromethorphan, and diphenhydramine spiked at 75%, 100%, and
150% concentration levels using USP Residual Solvents Class 2—Mixture C RS. For the accuracy and recovery of selected components of USP
Residual Solvents Class 2—Mixture A RS and USP Residual Solvents Class 2—Mixture B RS, we used 10-mg/mL solutions of acetaminophen,
aspirin, and ibuprofen spiked at 50%, 100%, and 200% concentration levels, respectively. For the accuracy and recovery for the remaining two
components of USP Residual Solvents Class 2—Mixture C RS (formamide and sulfolane), we used 40-mg/mL solutions of acetaminophen,
dextromethorphan, and diphenhydramine spiked at 50%, 100%, and 200% concentration levels, respectively.
Recovery data are summarized in Table 7 and Table 9. Specifically, we calculated the average recovery for three replicates at three
concentration levels. The acceptance criterion recommended in 〈1467〉 for nonchromatographic procedures is a solvent recovery within the 80%–
120% range. For most of these solvents in the specified substances, the recovery acceptance criteria were met. The exceptions are
dimethylacetamide and ethylene glycol in diphenhydramine; sulfolane in acetaminophen and diphenhydramine; tetrahydrofuran in aspirin and
ibuprofen; nitromethane in ibuprofen; and pyridine in acetaminophen, aspirin, and ibuprofen. Except for pyridine, the variability of recovery data
for these analytes is believed to be attributed to the valve performance. The variation in pyridine recovery data strongly suggests that matrix
effects may be responsible. However, because of both high MRR sensitivity to pyridine and a good linear dependence for the pyridine signal in
different substances, we believe that the standard addition method, similar to that described in 〈467〉, 8. Analytical Procedures for Class 1 and
Class 2 Residual Solvents, 8.1 Chromatographic Systems, Procedure C will produce accurate results for pyridine quantitation in pharmaceutical
substances despite the observed matrix effects.
Solution Stability
Fresh Standard stock solution and Standard solutions were prepared daily; therefore, the solution stability was not evaluated for this Stimuli
article.
Click image to enlarge
Figure 5. MRR validation data for solvents from USP Residual Solvents Class 2—Mixture C RS. For the top six solvents (2-methoxyethanol, 2-
ethoxyethanol, dimethylacetamide, dimethylformamide, ethylene glycol, and N-methylpyrrolidone), the corresponding standards were prepared
using USP Residual Solvents Class 2—Mixture C RS (see previous sections and Table 3). For formamide and sulfolane, the custom standards were
made as described previously and in Table 4 and Table 5. The regression line is provided for “no substance” data.
Table 6. Validation Data for Eight Solvents from USP Residual Solvents Class 2—Mixture C RS in Ionic Liquids (BMIMBF )
Solvent
Concentration
Range in
Solution
(µg/mL)
s
USP
Concentration
Limit (µg/g)
Linearity,
R2
Repeatability
(3 × 3, %)
MRR
Response
Factor
(µV per 1
µg/mL)
600s
Background
Noise (SD of 6
Measurements,
µV
600 s
LOQ in
Solution
(µg/mL)
Estimated 600 s LOQ
with Respect to
Substance (µg/g) for
Concentrations of:
10
mg/mL
20
mg/mL
50
mg/mL
2-Methoxyethanola 0.5–2.5 0.995 14.0 21.1 0.66 0.31 31 16 6 50
2-Ethoxyethanola 1.6–8 0.992 19.0 5.4 0.65 1.21 121 61 24 160
Dimethylformamide
(DMF)a 8.8–44 0.981 15.4 14.0 1.11 0.79 79 40 16 880
4
Solvent
Concentration
Range in
Solution
(µg/mL)
s
USP
Concentration
Limit (µg/g)
Linearity,
R
2
Repeatability
(3 × 3, %)
MRR
Response
Factor
(µV per 1
µg/mL)
600s
Background
Noise (SD of 6
Measurements,
µV
600 s
LOQ in
Solution
(µg/mL)
Estimated 600 s LOQ
with Respect to
Substance (µg/g) for
Concentrations of:
10
mg/mL
20
mg/mL
50
mg/mL
Dimethylacetamide
(DMAC)a 10.9–54.5 0.981 30.6 0.77 0.41 5.37 537 268 107 1090
Ethylene glycola 6.2–31 0.990 16.6 4.3 1.16 2.70 270 135 54 620
NMethylpyrrolidone
(NMP)a 5.3–26.5 0.969 15.7 1.9 0.59 3.06 306 153 61 530
Formamideb 4.4–35.2 0.980 10.0 1.4 0.70 5.11 511 255 102 220
Sulfolaneb 3.2–25.6 0.984 15.3 2.8 0.62 2.25 225 113 45 160
a USP Residual Solvents Class 2—Mixture C RS was used.
b Commercial individual neat solvents were used.
Table 7. Mean Recovery Data for Eight Solvents from USP Residual Solvents Class 2—Mixture C RS in 20-mg/mL Drug Substance
Solutions in Ionic Liquids (BMIMBF )
Solvent
Mean % Recovery (3 × 3) for 75%, 100%, and 150% Levels for:
Acetaminophen Dextromethorphan Diphenhydramine
2-Methoxyethanola 106.7 106.9 106.4
2-Ethoxyethanola 117.6 95.0 103.5
Dimethylformamide (DMF)a 114.3 99.9 100.8
Dimethylacetamide (DMAC)a 117.7 105.3 126.8
Ethylene glycola 118.6 90.3 64.3
N-Methylpyrrolidone (NMP)a 105.3 91.8 80.8
Formamideb 100.4 76.3 104.3
Sulfolaneb 78.8 107.5 146.5
a USP Residual Solvents Class 2—Mixture C RS used.
b Commercial individual neat solvents were used.
4
Click image to enlarge
Figure 6. Validation data for selected solvents of high and medium volatility from USP Residual Solvents Class 2—Mixture A RS and USP
Residual Solvents Class 2—Mixture B RS. Custom standards were prepared using pure residual solvents, as described previously and in Table 4
and Table 5.
Table 8. Validation Data for Selected Solvents from USP Residual Solvents Class 2—Mixture A RS and USP Residual Solvents
Class 2—Mixture B RS in Ionic Liquids (BMIMBF )
Solventa
Concentration
Range in
Solution
(µg/mL)
Evaluated Performance Parameter
USP Limit
in
Substance
(µg/g)
Linearity,
R
2
Repeatability
(3 × 3, %)
MRR
Response
Factor
(µV per 1
µg/mL)
600 s
Background
Noise (µV, SD
of 6
Measurements)
600 s
LOQ in
Solution
(µg/mL)
Estimated 600 s LOQ in
Substance (µg/g) for
Solution Concentration
Min. Max.
10
mg/mL
20
mg/mL
50
mg/mL
cis-1,2-
Dichloroethene 1.65 93.5 0.999 10.9 4.36 0.72 1.65 165 83 33 1870
4
Solventa
Concentration
Range in
Solution
(µg/mL)
Evaluated Performance Parameter
USP Limit
in
Substance
(µg/g)
Linearity,
R2
Repeatability
(3 × 3, %)
MRR
Response
Factor
(µV per 1
µg/mL)
600 s
Background
Noise (µV, SD
of 6
Measurements)
600 s
LOQ in
Solution
(µg/mL)
Estimated 600 s LOQ in
Substance (µg/g) for
Solution Concentration
Min. Max.
10
mg/mL
20
mg/mL
50
mg/mL
Methanol 1.99 150 0.997 6.1 4.43 0.88 1.99 199 99 40 3000
Tetrahydrofuran 5.75 36 0.996 25.6 1.74 1.00 5.75 575 287 115 720
Acetonitrile 0.09 20.5 0.994 11.9 72.3 0.63 0.09 8.7 4.4 1.7 410
Nitromethane 0.03 2.5 0.999 9.2 248 0.66 0.03 2.7 1.3 0.53 50
Toluene 3.15 44.5 0.994 13.8 3.17 1.00 3.15 315 158 63 890
Pyridine 0.38 10 0.997 8.9 12.7 0.48 0.38 38 19 8 200
Chlorobenzene 1.52 18 0.994 16.6 11.9 1.81 1.52 152 76 30 360
Ethylbenzene 1.77 18.4 0.989 25.3 4.47 0.79 1.77 177 88 35 368
m-Xylene 6.10 65.1 0.986 34.6 1.05 0.64 6.10 610 305 122 1302
a Neat commercial solvents were used (see Table 2).
Table 9. Recovery Data for Selected Solvents from USP Residual Solvents Class 2—Mixture A RS and USP Residual Solvents Class
2—Mixture B RS in Drug Substance Solutions with a Concentration of 10 mg/mL
Residual Solvent
Average Recovery (%) for 50%, 100%, and 200% Concentration Levelsa
Acetaminophen Aspirin Ibuprofen
cis-1,2-Dichloroethene 93. 5 110.1 95.2
Methanol 97.3 91.5 92.4
Tetrahydrofuran 110.0 121.7 121.0
Acetonitrile 101.9 107.5 104.8
Nitromethane (2-B) 86.1 82.4 79.4
Toluene (2-A) 86.8 95.8 96.2
Pyridine (2-B) 65.0 52.5 79.1
a Three preparations per level.
Click image to enlarge
Figure 7. Validation data for selected solvents of medium volatility from USP Residual Solvents Class 2—Mixture A RS: no substance (black dot)
vs. 10-mg/mL acetaminophen (red x). Additional studies are being conducted. Custom residual solvent standards were prepared as described
previously and in Table 4 and Table 5.
Discussion
In contrast to chromatography, which determines different component compounds of a mixture based on chromatographic separation (i.e.,
temporal separation because of the distribution of components between two phases), the headspace-MRR identifies and quantifies those
components directly in the gas phase based on their pure molecular rotational spectrum resulting from molecular geometry and three-
dimensional mass distributions. All volatile components of a sample are present in the sample gas phase during MRR analysis. To ensure linear
response of MRR signals, the total vapor pressure of all volatile components (but not counting the carrier gas) generally should not exceed 1 Torr.
This can be achieved by adjusting the headspace temperature and/or changing the composition of a sample matrix (for example, by changing the
sample preparation procedure). Therefore, it is beneficial for MRR analysis to use diluents with as low as possible vapor pressure. For this reason,
we used ionic liquids or ionic liquids with small amounts of low-volatile cosolvents (such as sulfolane).
One more important consideration that can help to significantly improve the sensitivity and speed of MRR analysis is to reconsider not only the
diluent but also the grouping of the residual solvents in standard mixtures. Specifically, the commercially available standard mixtures for residual
solvent analysis, such as USP Residual Solvents Class 2—Mixture A RS, USP Residual Solvents Class 2—Mixture B RS, and USP Residual Solvents
Class 2—Mixture C RS, are designed for GC. In contrast, to optimize MRR analytical metrics, it would be beneficial to make standard mixtures of
residual solvents that are grouped by their boiling points and avoid placing highly volatile and low-volatile solvents into one standard mixture. In
this work, we proved the feasibility of this approach using custom-made standards for selected residual solvents from USP Residual Solvents
Class 2—Mixture A RS, USP Residual Solvents Class 2—Mixture B RS, and USP Residual Solvents Class 2—Mixture C RS (Table 4 and Table 5).
Despite the above, USP Residual Solvents Class 2—Mixture C RS as described in 〈467〉 can certainly be used with MRR to directly analyze six (of
eight) solvents represented in the Reference Standard: 2-methoxyethanol, 2-etoxyethanol, dimethylformamide, dimethyl acetamide, ethylene
glycol, and N-methylpyrrolidone; thus filling a gap in the analysis for these solvents as there is no compendial or any other standard procedure
available. The same conclusion can likely be derived for many solvents represented in USP Residual Solvents Class 2—Mixture A RS and USP
Residual Solvents Class 2—Mixture B RS, but the exact list of solvents suitable for the conventional workflow using commercial standard mixtures
is still to be determined.
In contrast, if custom standards that utilize low-volatile diluent and residual solvents grouped by their boiling points are made, MRR analysis is
certainly capable of meeting analytical technique performance characteristics needed for complying with ICH and USP acceptance criteria (per
〈467〉) for most Class 2 solvents. The exceptions include nonpolar solvents such as trans-1,2-dichloroethene; 1,4-dioxane;p-xylene; and nhexane.
CONCLUSIONS
This Stimuli article demonstrates that MRR spectroscopy can emerge as an alternative method for residual solvent analysis in the near future.
This is due to several factors. First, MRR bridges the analytical gaps in conventional SH-GC methods for analyzing low-volatile residual solvents,
such as those from USP Residual Solvents Class 2—Mixture C RS. Second, the high chemical selectivity of MRR enables direct analysis of residual
solvents in multicomponent mixtures without chemical separation to significantly decrease analysis time comparing to GC and eliminate any
potential analyte coelution and identification problems.
The developed MRR method was validated for several selected residual solvents from USP Residual Solvents Class 2—Mixture A RS and USP
Residual Solvents Class 2—Mixture B RS (see 〈467〉), as well as for all eight low-volatile solvents from USP Residual Solvents Class 2—Mixture C
RS. The first-generation headspace-MRR prototype that was used in this study is capable of meeting analytical procedure performance
characteristics for complying with USP acceptance criteria (per 〈467〉) for most of the studied residual solvents. In addition, the first tests of a
new-generation fully-automated headspace-MRR setup show significantly improved analysis performance characteristics in terms of analysis
sensitivity, repeatability, and sample throughput. With simultaneous multi-sample headspace equilibration utilized in this new setup, it is possible
to achieve effective residual solvent analysis cycle times between 2 and 20 min per sample, depending on the number and type of analytes and
required analysis sensitivity.
In summary, as the sensitivity of MRR analysis can be accurately predicted theoretically, it is possible to state with high confidence that the
new-generation, fully-automated, high-temperature headspace-MRR instrument can meet analytical procedure performance characteristics
needed for complying with USP acceptance criteria (per 〈467〉) for most of Class 2 and 3 solvents including those from USP Residual Solvents
Class 2—Mixture C RS, water-soluble acids, and DMSO, as well as about half of Class 1 solvents. In addition, MRR can directly analyze several
other volatile impurities poorly suited to GC analysis that are not listed in 〈467〉. This list includes but is not limited to formaldehyde, ethylene
oxide, carboxylic acids, and volatile amines.
Furthermore, MRR is an online-capable technique (13,21) that can bring the high-resolution and extraordinary chemical selectivity of high-end
instrumentation directly to the process floor. For this reason, MRR can support the implementation of continuous manufacturing, as well as PAT
and QbD initiatives to enable detection of diverse impurities at any stage of pharmaceutical development and manufacturing.
APPENDIX: THEORETICAL ESTIMATE OF TOTAL VAPOR PRESSURE OF ALL SAMPLE COMPONENTS IN HEADSPACE
For each component of the sample mixture in the sealed headspace vial, including the diluent, the equilibrium headspace concentration of that
component in the neon gas that exits the headspace vial during MRR measurement, is directly proportional to its partial vapor pressure,
Equation 1
According to Raoult’s law (53), the equilibrium vapor pressure of i
th
mixture component, is proportional to its molar fraction in solution,
. In ideal mixtures, the proportionality constant is equal to the vapor pressure of the pure analyte, , at a given
headspace temperature T.
Equation 2
c
gas
i
,
p
gas
i
:
c
gas
i ∝ p
gas
i
p
gas
i
,
μ
solution
i p
100%
i (T)
p
gas
i = p
100%
i (T) ∙ μ
solution
i
In real mixtures, there are deviations from Raoult’s law. The matrix effects mainly occur because of the intermolecular interactions in a liquid
phase between the i
th
analyte and the other matrix components, which may change the proportionality coefficient between the partial vapor
pressure of a component and its molar fraction in solution. To compensate for these effects, the activity coefficient, , can be introduced to
Raoult's equation (53):
Equation 3
According to Dalton’s law (53), total vapor pressure of components of a gas-phase mixture, , can be determined as a sum of partial
vapor pressures, , of individual components of a mixture:
Equation 4
In their turn, vapor pressures of pure analytes, , can be modeled using Antoine’s equation (54):
Equation 5
where T is the temperature of the solution and A , B , and C are the Antoine’s equation coefficients for the i
th
component of a mixture. It should
be noted that Antoine's coefficients are tabulated for many substances, including residual solvents, and can be found in the literature (55–56).
By substituting Equation 5 into Equation 4, it is possible to derive the equation that is used for estimating the total vapor pressure of volatile
components of any given sample mixture and at any given headspace temperature:
Equation 6
It should be noted that for dilute solutions, it is possible to introduce the direct proportionality coefficient that reflects the equilibrium
distribution between the gas-phase and solution-phase molar fractions of a particular analyte, and , respectively. This
coefficient, is often called the partition coefficient (or, sometimes, the distribution coefficient) (53). The partition coefficient is
temperature-dependent but constant at any given temperature. From comparison to Equation 4, it is obvious that this coefficient is reversely
proportional to the vapor pressure of the corresponding analyte:
Equation 7
With the aid of partition coefficients, Equation 4 can be redefined to enable determination of the total molar concentration of all components in
the gas phase from their corresponding molar concentrations in the solution phase:
Equation 8
It is evident from Equation 8 that volatile analytes, for example, most solvents from USP Residual Solvents Class 2—Mixture A RS and USP
Residual Solvents Class 2—Mixture B RS, will have small values of K and, thus, much higher molar fractions in a gas phase compared to those in
a liquid phase. In contrast, nonvolatile sample components, such as the drug substance and excipients, will have very large values of Ki (i.e.,
very low vapor pressures) and, therefore, will not contribute to the gas phase composition.
In this article, we consider all xylene isomers (o-xylene, m-xylene, p-xylene, and ethylbenzene) and the two 1,2-dichloroethene isomers (cis
and trans) to be separate species, as they have distinct MRR spectra.
REFERENCES
1. ICH Q2(R2): Validation of Analytical Procedures. International Council for Harmonization: 2022; pp 1–34.
2. Residual Solvents: Alternate Validated Methods; 2012. cdnmedia.eurofins.com/eurofins-us/media/447610/9405_residual_solvents.pdf.
3. Analysis of USP Residual Solvents of Class 1, Class 2, and Class 3 using the Agilent 8890 GC/FID/5977B MSD System;
www.agilent.com/cs/library/applications/application-usp467-residual%20solvents-class1-2-3-5994-1488en-agilent.pdf.
4. Witschi C, Doelker E. Residual solvents in pharmaceutical products: acceptable limits, influences on physicochemical properties, analytical
methods and documented values. European Journal of Pharmaceutics and Biopharmaceutics. 1997;43(3):215–242. DOI: 10.1016/S0939-
6411(96)00037-9.
5. Smith ID, Waters DG. Determination of residual solvent levels in bulk pharmaceuticals by capillary gas chromatography. Analyst.
1991;116(12):1327–1331. DOI: 10.1039/AN9911601327.
6. B'Hymer C. Residual solvent testing: a review of gas-chromatographic and alternative techniques. Pharmaceutical Research.
2003;20(3):337–344. DOI: 10.1023/A:1022693516409.
7. Shea JA, Cook SC, Shamrock WF, Segretario J. Quantitation of residual N-methylpyrrolidone in losoxantrone hydrochloride by reversedphase high-performance liquid chromatography. Journal of Chromatographic Science. 1998;36(4):187–190. DOI:
10.1093/chromsci/36.4.187.
γi
p
gas
i = p
100%
i (T) ∙ γi ∙ μ
solution
i
p
gas
total
p
gas
i
p
gas
total = ∑(p
gas
i ) = ∑[p
100%
i (T) ∙ γi ∙ μ
solution
i ]
p
100%
i (T)
log10
[p
100%
i (T)] = Ai − Bi/ [T + Ci]
i i i
p
gas
total(T) = ∑[10
( Ai −
Bi
(T+Ci) ) ∙ γi ∙ μ
solution
i ]
μ
gas
i μ
solution
i
Ki(T),
Ki(T) = μ
solution
i /μ
gas
i ∝ 1/[p
100%
i (T) ∙ γi]
μ
gas
total = ∑i μ
gas
i = ∑i μ
solution
i /Ki(T)
i
8. Raghani AR. High-speed gas chromatographic analysis of solvents in pharmaceuticals using solid phase microextraction. Journal of
Pharmaceutical and Biomedical Analysis. 2002;29(3):507–518. DOI: 10.1016/S0731-7085(02)00085-7.
9. Camarasu CC. Unknown residual solvents identification in drug products by headspace solid-phase microextraction gas chromatographymass spectrometry. Chromatographia. 2002;56(1). DOI: 10.1007/BF02494126.
10. Biba EP, Langford VS. High-throughput residual solvent analysis using selected ion flow tube mass spectrometry (SIFT-MS). In: Pharm.
Forum. 2021.
11. Neill JL, Evangelisti L, Pate BH. Analysis of isomeric mixtures by molecular rotational resonance spectroscopy. Analytical Science Advances.
2023;4(5–6):204–219. DOI: 10.1002/ansa.202300021.
12. Neill JL, Mikhonin AV, Chen T, Sonstrom RE, Pate BH. Rapid quantification of isomeric and dehalogenated impurities in pharmaceutical raw
materials using MRR spectroscopy. Journal of Pharmaceutical and Biomedical Analysis. 2020;189:113474. DOI:
10.1016/j.jpba.2020.113474.
13. Neill JL, Yang Y, Muckle MT, Reynolds RL, Evangelisti L, Sonstrom RE, Pate BH, Gupton BF. Online stereochemical process monitoring by
molecular rotational resonance spectroscopy. Org Process Res Dev. 2019;23(5):1046–1051.
14. Armstrong DW, Talebi M, Thakur N, Wahab MF, Mikhonin AV, Muckle MT, Neill JL. A gas chromatography-molecular rotational resonance
spectroscopy-based system of singular specificity. Angewandte Chemie International Edition. 2020;59(1):192–196. DOI:
10.1002/anie.201910507.
15. Vang ZP, Reyes A, Sonstrom RE, Holdren MS, Sloane SE, Alansari IY, Neill JL, Pate BH, Clark JR. Copper-catalyzed transfer hydrodeuteration
of aryl alkenes with quantitative isotopomer purity analysis by molecular rotational resonance spectroscopy. Journal of the American
Chemical Society. 2021;143(20):7707–7718. DOI: 10.1021/jacs.1c00884.
16. Sonstrom RE, Neill JL, Mikhonin AV, Doetzer R, Pate BH. Chiral analysis of pantolactone with molecular rotational resonance spectroscopy.
Chirality. 2022;34(1):114–125. DOI: 10.1002/chir.23379.
17. Sonstrom RE, Cannon DM, Neill JL. Chiral analysis of linalool, an important natural fragrance and flavor compound, by molecular rotational
resonance spectroscopy. Symmetry. 2022;14(5):917.
18. Mills MD, Sonstrom RE, Vang ZP, Neill JL, Scolati HN, West CT, Pate BH, Clark JR. Enantioselective synthesis of enantioisotopomers with
quantitative chiral analysis by chiral tag rotational spectroscopy. Angewandte Chemie International Edition. 2022;61(33). DOI:
10.1002/anie.202207275.
19. Sonstrom RE, Vang ZP, Scolati HN, Neill JL, Pate BH, Clark JR. Rapid enantiomeric excess measurements of enantioisotopomers by
molecular rotational resonance spectroscopy. Organic Process Research & Development. 2023;27(7):1185–1197. DOI:
10.1021/acs.oprd.3c00028.
20. Joyce LA, Schultz DM, Sherer EC, Neill JL, Sonstrom RE, Pate BH. Direct regioisomer analysis of crude reaction mixtures via molecular
rotational resonance (MRR) spectroscopy. Chemical Science. 2020;11(24):6332–6338. DOI: 10.1039/D0SC01853H.
21. Byars AA, Kompally KR, Mechnick E, Sonstrom RE, Mikhonin A, Neill JL, Boetzel R, MacGregor J, Talicska C, Li J, et al. An automated, highly
selective reaction monitoring instrument using molecular rotational resonance spectroscopy. Precision Chemistry. 2024;2(2):57–62. DOI:
10.1021/prechem.3c00098.
22. Legon AC. Pulsed-nozzle, Fourier-transform microwave spectroscopy of weakly bound dimers. Annual Review of Physical Chemistry.
1983;34:275–300. DOI: 10.1146/annurev.pc.34.100183.001423.
23. Brown GG, Dian BC, Douglass KO, Geyer SM, Shipman ST, Pate BH. A broadband Fourier transform microwave spectrometer based on
chirped pulse excitation. Rev Sci Instrum. 2008;79(5):053103. DOI: 10.1063/1.2919120.
24. Grimme S, Hansen A, Brandenburg JG, Bannwarth C. Dispersion-corrected mean-field electronic structure methods. Chemical Reviews.
2016;116(9):5105–5154. DOI: 10.1021/acs.chemrev.5b00533.
25. Buckley P, Brochu M. Microwave spectrum, dipole moment, and intramolecular hydrogen bond of 2-methoxyethanol. Canadian Journal of
Chemistry. 1972;50(8):1149–1156. DOI: 10.1139/v72-182.
26. Phillips MA, Shipman ST. Waveguide chirped-pulse fourier transform microwave spectroscopy of 2-ethoxyethanol. In: 68th International
Symposium on Molecular Spectroscopy, June 01, 2013, 2013; p ERK05.
27. Fujitake M, Kubota Y, Ohashi N. Fourier transform microwave spectroscopy of N,N-dimethylacetamide. Journal of Molecular Spectroscopy.
2006;236(1):97–109. DOI: 10.1016/j.jms.2005.12.013.
28. Heineking N, Dreizler H. Nuclear quadrupole hyperfine structure and methyl torsional fine structure in the rotational spectra of N,Ndimethylformamide and N-nitrosodimethylamine. Zeitschrift für Naturforschung A. 1993;48(4):570–576. DOI: 10.1515/zna-1993-0406.
29. Melosso M, Dore L, Tamassia F, Brogan CL, Hunter TR, McGuire BA. The submillimeter rotational spectrum of ethylene glycol up to 890 GHz
and application to ALMA band 10 spectral line data of NGC 6334I. The Journal of Physical Chemistry A. 2020;124(1):240–246. DOI:
10.1021/acs.jpca.9b10803.
30. Hirota E, Sugisaki R, Nielsen CJ, Sørensen GO. Molecular structure and internal motion of formamide from microwave spectrum. Journal of
Molecular Spectroscopy. 1974;49(2):251–267. DOI: 10.1016/0022-2852(74)90274-4.
31. Alonso JL. Microwave spectrum of sulfolane. Journal of Molecular Structure. 1981;73(1):71–78. DOI: 10.1016/0022-2860(81)85047-8.
32. Boucher D, Burie J, Bauer A, Dubrulle A, Demaison J. Microwave spectra of molecules of astrophysical interest. XIX. methyl cyanide.
Journal of Physical and Chemical Reference Data. 1980;9(3):659–720. DOI: 10.1063/1.555625.
33. Dorosh O, Białkowska-Jaworska E, Kisiel Z, Pszczółkowski L. New measurements and global analysis of rotational spectra of Cl-, Br-, and Ibenzene: Spectroscopic constants and electric dipole moments. Journal of Molecular Spectroscopy. 2007;246(2):228–232. DOI:
10.1016/j.jms.2007.09.010.
34. Flygare WH, Howe JA. Microwave spectrum and quadrupole interaction in cis‐1,2‐dichloroethylene. The Journal of Chemical Physics.
1962;36(2):440–443. DOI: 10.1063/1.1732527.
35. Maté B, Suenram RD, Lugez C. Microwave studies of three alkylbenzenes: Ethyl, n-propyl, and n-butylbenzene. The Journal of Chemical
Physics. 2000;113(1):192–199. DOI: 10.1063/1.481818.
36. True NS, Farag MS, Bohn RK, MacGregor MA, Radhakrishnan J. Low-resolution microwave studies of substituted ethyl- and
isopropylbenzenes. The Journal of Physical Chemistry. 1983;87(23):4622–4627. DOI: 10.1021/j100246a018.
37. Xu LH, Lovas FJ. Microwave spectra of molecules of astrophysical interest. XXIV. Methanol (CH OH and 13CH OH). Journal of Physical and
Chemical Reference Data. 1997;26(1):17–156. DOI: 10.1063/1.556005.
38. Zhao Y, Stahl W, Nguyen HVL. Ketone physics—structure, conformations, and dynamics of methyl isobutyl ketone explored by microwave
spectroscopy and quantum chemical calculations. Chemical Physics Letters. 2012;545:9–13. DOI: 10.1016/j.cplett.2012.07.009.
39. Kisiel Z, Kosarzewski J, Pszczółkowski L. Nuclear quadrupole coupling tensor of CH2Cl2: comparison of quadrupolar and structural angles in
methylene halides. Acta Phys Pol A. 1997;92(3):507–516. DOI: 10.12693/APhysPolA.92.507.
40. Mamleev AH, Gunderova LN, Galeev RV. Microwave spectrum and hindered pseudorotation of tetrahydrofuran. Journal of Structural
Chemistry. 2001;42(3):365–370. DOI: 10.1023/A:1012448618502.
41. Ilyushin VV, Kisiel Z, Pszczółkowski L, Mäder H, Hougen JT. A new torsion–rotation fitting program for molecules with a sixfold barrier:
Application to the microwave spectrum of toluene. Journal of Molecular Spectroscopy. 2010;259(1):26–38. DOI:
10.1016/j.jms.2009.10.005.
42. Thomsen C, Dreizler H. The microwave spectra of m-xylene and m-xylene-d10. Determination of the low methyl internal rotation barrier.
2001;56(9–10):635–640. DOI: doi:10.1515/zna-2001-0905.
43. Rudolph HD, Walzer K, Krutzik I. Microwave spectrum, barrier for methyl rotation, methyl conformation, and dipole moment of orthoxylene. Journal of Molecular Spectroscopy. 1973;47(2):314–339. DOI: 10.1016/0022-2852(73)90016-7.
44. Carpenter JH, Seo PJ, Whiffen DH. The rotational spectrum of chloroform in its ground and excited vibrational states. Journal of Molecular
Spectroscopy. 1995;170(1):215–227. DOI: 10.1006/jmsp.1995.1066.
45. Li W, Vigorito A, Calabrese C, Evangelisti L, Favero LB, Maris A, Melandri S. The microwave spectroscopy study of 1,2-dimethoxyethane.
Journal of Molecular Spectroscopy. 2017;337:3–8. DOI: 10.1016/j.jms.2017.02.015.
46. Fritz SM, Mishra P, Zwier TS. Strong-field coherence breaking as a tool for identifying methyl rotor states in microwave spectra: 2-
hexanone. The Journal of Chemical Physics. 2019;151(4):041104. DOI: 10.1063/1.5109656.
47. Changala PB, Baraban JH. Ab initio effective rotational and rovibrational Hamiltonians for non-rigid systems via curvilinear second order
vibrational Møller–Plesset perturbation theory. The Journal of Chemical Physics. 2016;145(17):174106. DOI: 10.1063/1.4966234.
48. Rohart F. Microwave spectrum of nitromethane internal rotation Hamiltonian in the low barrier case. Journal of Molecular Spectroscopy.
1975;57(2):301–311. DOI: 10.1016/0022-2852(75)90033-8.
49. DeMore BB, Wilcox WS, Goldstein JH. Microwave spectrum and dipole moment of pyridine. The Journal of Chemical Physics.
1954;22(5):876–877. DOI: 10.1063/1.1740205.
50. Maris A, Favero LB, Danieli R, Favero PG, Caminati W. Vibrational relaxation in pyridine upon supersonic expansion. The Journal of Chemical
Physics. 2000;113(19):8567–8573. DOI: 10.1063/1.1318224.
51. Kisiel Z, Pszczółkowski L. Assignment and analysis of the MM-Wave rotational spectrum of trichloroethylene: Observation of a new,
extended bR-band and an overview of high-J,R-type bands. Journal of Molecular Spectroscopy. 1996;178(1):125–137. DOI:
10.1006/jmsp.1996.0165.
52. Taokaew S, Kriangkrai W. Recent progress in processing cellulose using ionic liquids as solvents. Polysaccharides, 2022; Vol. 3, pp 671–
691.
53. Kolb BE, Leslie K. Static Headspace-Gas Chromatography: Theory and Practice; John Wiley & Sons, Inc., Hoboken, NJ, 2006.
54. Wisniak J. Historical development of the vapor pressure equation from Dalton to Antoine. Journal of Phase Equilibria. 2001;22(6):622–630.
DOI: 10.1007/s11669-001-0026-x.
55. NIST Chemistry WebBook, SRD 69; National Institute of Standards and Technology.
56. Yaws CL. The Yaws Handbook of Vapor Pressure: Antoine Coefficients; Gulf Professional Publishing is an imprint of Elsevier, 2015.
a BrightSpec, Inc., 770 Harris St., Charlottesville, VA, United States.
b United States Pharmacopeia, 12601 Twinbrook Parkway, Rockville, MD, United States.
Page Information:
Not Applicable
DocID:
3 3